Clinical Performance of a Novel 15.5-Day Factory-Calibrated Continuous Glucose Monitoring System in Adults With Diabetes
BACKGROUND: Performance of the uCGM 100 continuous glucose monitoring (CGM) system, featuring a 15.5-day, real-time, and factory-calibrated glucose sensor, is evaluated against venous plasma reference for adult participants (≥18 years) in a multicenter, prospective clinical study. METHODS: A total number of 60 participants were enrolled at three clinical sites in China, and each participant wore sensors on the back of each upper arm for 15.5 days. Three in-clinic visits, lasting eight hours during each visit, were conducted by each participant to perform multiple venous plasma glucose measurements using Biosen C-Line GP+ (EKF Diagnostics) Glucose/Lactate Measuring System. Venous plasma blood glucose sample tests covered the start, middle, and end wear of the sensor (day 1, days 3-5, days 7-9, days 11-13, or the last 24 hours). Analytic performance included accuracy such as ±20% or ±20 mg/dL (%20/20) agreement rate with reference values, and mean absolute relative difference (MARD) between CGM and reference values. RESULTS: A total of 11 576 matched data pairs (120 sensors) were analyzed. The overall MARD and %20/20 agreement rate for the CGM system was 7.6% and 95.5%, respectively, compared with venous blood reference. The %20/20 agreement remained consistently above 93% across wear days and glucose ranges. No serious adverse events were recorded. Usability survey showed 99.8% positive feedback, demonstrating good usability. CONCLUSIONS: The novel CGM system was proven to provide accurate and reliable glucose readings in adults with diabetes.
Authors
- X.C. Li (ORCID: https://orcid.org/0000-0002-5981-2762)
- Yanrong Yang (ORCID: https://orcid.org/0000-0002-2642-2207)
- Hongying Ye (ORCID: https://orcid.org/0000-0002-6932-3145)
- Yang Li (ORCID: https://orcid.org/0000-0003-3387-1913)
- Tikuan Cai
- Lin Zhao
- Bo Feng
Institutions
- Sun Yat-sen University (CN)
- Fudan University (CN)
- Zhongshan Hospital (CN)
- Shanghai East Hospital (CN)
- The First Affiliated Hospital, Sun Yat-sen University (CN)
- Shanghai Fudan Microelectronics (China) (CN)
- United Imaging Healthcare (China) (CN)
- Huashan Hospital (CN)
- United Microelectronics (United States) (US)
Publication Details
- Journal
- Journal of Diabetes Science and Technology
- Published
- 2026-05-18
- DOI
- https://doi.org/10.1177/19322968261447295
- Primary Topic
- Diabetes Management and Research
- Type
- article
- Field-Weighted Citation Impact
- 0.00