Clinical Performance of a Novel 15.5-Day Factory-Calibrated Continuous Glucose Monitoring System in Adults With Diabetes

BACKGROUND: Performance of the uCGM 100 continuous glucose monitoring (CGM) system, featuring a 15.5-day, real-time, and factory-calibrated glucose sensor, is evaluated against venous plasma reference for adult participants (≥18 years) in a multicenter, prospective clinical study. METHODS: A total number of 60 participants were enrolled at three clinical sites in China, and each participant wore sensors on the back of each upper arm for 15.5 days. Three in-clinic visits, lasting eight hours during each visit, were conducted by each participant to perform multiple venous plasma glucose measurements using Biosen C-Line GP+ (EKF Diagnostics) Glucose/Lactate Measuring System. Venous plasma blood glucose sample tests covered the start, middle, and end wear of the sensor (day 1, days 3-5, days 7-9, days 11-13, or the last 24 hours). Analytic performance included accuracy such as ±20% or ±20 mg/dL (%20/20) agreement rate with reference values, and mean absolute relative difference (MARD) between CGM and reference values. RESULTS: A total of 11 576 matched data pairs (120 sensors) were analyzed. The overall MARD and %20/20 agreement rate for the CGM system was 7.6% and 95.5%, respectively, compared with venous blood reference. The %20/20 agreement remained consistently above 93% across wear days and glucose ranges. No serious adverse events were recorded. Usability survey showed 99.8% positive feedback, demonstrating good usability. CONCLUSIONS: The novel CGM system was proven to provide accurate and reliable glucose readings in adults with diabetes.

Authors

Institutions

Publication Details

Journal
Journal of Diabetes Science and Technology
Published
2026-05-18
DOI
https://doi.org/10.1177/19322968261447295
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Performance of a Novel 15.5-Day Factory-Calibrated Continuous Glucose Monitoring System in Adults With Diabetes

X.C. Li, Yanrong Yang, Hongying Ye, Yang Li et al.
Journal of Diabetes Science and Technology
Diabetes Management and Research
article

Clinical Performance of a Novel 15.5-Day Factory-Calibrated Continuous Glucose Monitoring System in Adults With Diabetes

X.C. Li, Yanrong Yang, Hongying Ye, Yang Li, Tikuan Cai, Lin Zhao, Bo Feng
article en

Abstract

BACKGROUND: Performance of the uCGM 100 continuous glucose monitoring (CGM) system, featuring a 15.5-day, real-time, and factory-calibrated glucose sensor, is evaluated against venous plasma reference for adult participants (≥18 years) in a multicenter, prospective clinical study. METHODS: A total number of 60 participants were enrolled at three clinical sites in China, and each participant wore sensors on the back of each upper arm for 15.5 days. Three in-clinic visits, lasting eight hours during each visit, were conducted by each participant to perform multiple venous plasma glucose measurements using Biosen C-Line GP+ (EKF Diagnostics) Glucose/Lactate Measuring System. Venous plasma blood glucose sample tests covered the start, middle, and end wear of the sensor (day 1, days 3-5, days 7-9, days 11-13, or the last 24 hours). Analytic performance included accuracy such as ±20% or ±20 mg/dL (%20/20) agreement rate with reference values, and mean absolute relative difference (MARD) between CGM and reference values. RESULTS: A total of 11 576 matched data pairs (120 sensors) were analyzed. The overall MARD and %20/20 agreement rate for the CGM system was 7.6% and 95.5%, respectively, compared with venous blood reference. The %20/20 agreement remained consistently above 93% across wear days and glucose ranges. No serious adverse events were recorded. Usability survey showed 99.8% positive feedback, demonstrating good usability. CONCLUSIONS: The novel CGM system was proven to provide accurate and reliable glucose readings in adults with diabetes.

Journal of Diabetes Science and Technology
Sun Yat-sen University (CN), Fudan University (CN), Zhongshan Hospital (CN), Shanghai East Hospital (CN), The First Affiliated Hospital, Sun Yat-sen University (CN), Shanghai Fudan Microelectronics (China) (CN), United Imaging Healthcare (China) (CN), Huashan Hospital (CN), United Microelectronics (United States) (US)
Quality Education
Openalex Percentile: Top 6%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.