Biomechanical comparison of two-screw and four-screw zero-profile fusion devices in anterior cervical discectomy and fusion using finite element analysis
Both two-screw and four-screw zero-profile fusion devices are commonly used in anterior cervical discectomy and fusion (ACDF), yet their biomechanical differences remain unclear. This study compared their biomechanical performance using a validated three-dimensional finite element (FE) model of the C2-T1 cervical spine reconstructed from computed tomography data. The devices were implanted at the C5/6 segment to simulate ACDF under various motion conditions. A 75 N follower load and a 1.0 N·m moment were applied to induce flexion, extension, lateral bending, and axial rotation. The two-screw zero-profile device demonstrated higher von Mises stress on the cage and plate but lower stress on the screws than the four-screw device. Stress distributions on the cage, screws, and adjacent vertebral bodies were comparable across models. After fusion, both devices exhibited similar stress characteristics. Overall, the two-screw zero-profile fusion device may achieve biomechanical behavior comparable to that of the four-screw zero-profile fusion device under the investigated static loading scenario. Further long-term clinical and biomechanical investigations are warranted to clarify its biomechanical behavior and potential implications following ACDF.
Authors
- Xingjin Wang (ORCID: https://orcid.org/0000-0001-5555-9109)
- H L Liu
- Kangkang Huang
- Wang Beiyu
- Chengyi Huang
- Tingkui Wu
- Minghe Yao
Institutions
- Heidelberg University (DE)
- Sichuan University (CN)
- University Hospital Heidelberg (DE)
- West China Hospital of Sichuan University (CN)
Publication Details
- Journal
- Scientific Reports
- Published
- 2026-05-18
- DOI
- https://doi.org/10.1038/s41598-026-51154-1
- Primary Topic
- Cervical and Thoracic Myelopathy
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Natural Science Foundation of Sichuan Province