Radiation shield for neonatal incubators: design and dosimetric evaluation in NICU radiography

Portable radiography in neonatal intensive care units (NICUs) produces scatter that can expose nearby neonates and healthcare workers, and space constraints may limit distance-based protection. This study aims to develop and validate a practical incubator-mounted shielding device and quantify its effect on scatter and projected occupational dose. A 0.5-mm Pb-equivalent sheet was shaped to fit the incubator, covered with cleanable vinyl, and secured with Velcro. Scatter was measured using a PMMA phantom with a 25.4 × 25.4 cm cross-section and 10-cm thickness at multiple angles and distances (0.5-2.0 m) for four exposure techniques (60-90 kVp) with and without shielding, including inside-incubator and adjacent-incubator scenarios. Angle-averaged dose at 1 m decreased by 92% (60 kVp/2 mAs), 86% (70 kVp/10 mAs), 85% (80 kVp/20 mAs) and 84% (90 kVp/63 mAs), with reductions consistent across angles and distances. Dual deployment further reduced scatter to adjacent incubators and nearby staff positions. Using institutional workload assumptions, projected annual occupational doses in the common-use 60 kVp/2 mAs scenario were already low at baseline and decreased by >90% with shielding, remaining <0.1 mSv/year at all tested distances. At 50 cm under this scenario, the estimate decreased from 0.034579 to 0.002871 mSv/year. User acceptance was surveyed among nurses and radiologic technologists (mean 4.59/5). This practical shielding approach provides consistent scatter reduction and may complement distance and exposure optimization in NICU radiography when space is limited.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-05-17
DOI
https://doi.org/10.1038/s41598-026-51681-x
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation shield for neonatal incubators: design and dosimetric evaluation in NICU radiography

Suphalak Khamruang Marshall, Sitthichok Chaichulee, Jongwat Cheewakul, Natee Ina et al.
Scientific Reports
Radiation Dose and Imaging
article

Radiation shield for neonatal incubators: design and dosimetric evaluation in NICU radiography

Suphalak Khamruang Marshall, Sitthichok Chaichulee, Jongwat Cheewakul, Natee Ina, Sumalee Aonprawat
article en

Abstract

Portable radiography in neonatal intensive care units (NICUs) produces scatter that can expose nearby neonates and healthcare workers, and space constraints may limit distance-based protection. This study aims to develop and validate a practical incubator-mounted shielding device and quantify its effect on scatter and projected occupational dose. A 0.5-mm Pb-equivalent sheet was shaped to fit the incubator, covered with cleanable vinyl, and secured with Velcro. Scatter was measured using a PMMA phantom with a 25.4 × 25.4 cm cross-section and 10-cm thickness at multiple angles and distances (0.5-2.0 m) for four exposure techniques (60-90 kVp) with and without shielding, including inside-incubator and adjacent-incubator scenarios. Angle-averaged dose at 1 m decreased by 92% (60 kVp/2 mAs), 86% (70 kVp/10 mAs), 85% (80 kVp/20 mAs) and 84% (90 kVp/63 mAs), with reductions consistent across angles and distances. Dual deployment further reduced scatter to adjacent incubators and nearby staff positions. Using institutional workload assumptions, projected annual occupational doses in the common-use 60 kVp/2 mAs scenario were already low at baseline and decreased by >90% with shielding, remaining <0.1 mSv/year at all tested distances. At 50 cm under this scenario, the estimate decreased from 0.034579 to 0.002871 mSv/year. User acceptance was surveyed among nurses and radiologic technologists (mean 4.59/5). This practical shielding approach provides consistent scatter reduction and may complement distance and exposure optimization in NICU radiography when space is limited.

Scientific Reports
Prince of Songkla University (TH), Songklanagarind Hospital (TH)
Openalex Percentile: Top 6%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Radiation shield for neonatal incubators: design and dosimetric evaluation in NICU radiography — Suphalak Khamruang Marshall, Sitthichok Chaichulee, et al. · Scientific Reports (2026) | TGRS Research Map | TGRS