A Comparative Evaluation of Two Rapid Strategies for Visual Field Testing: SITA Faster and SPARK Quick.

OBJECTIVES: This study compares the diagnostic performance of two rapid visual field testing strategies, SITA Faster (SF) and SPARK Quick (SQ), against the gold standard SITA Standard (SS). The measurement accuracy, agreement, and repeatability were evaluated. DESIGN: Prospective, cross-sectional study. PARTICIPANTS: A total of 118 eyes of 118 participants were enrolled. METHODS: Participants from a glaucoma clinic underwent multiple visual field tests. All subjects were naïve to SF and SQ, with SS 24-2 serving as the initial test. SF and SQ were then administered in random order. Among the 74 participants with visual field defects detected by SS, 49 underwent testing twice with both rapid strategies. Parameters were compared using one-way ANOVA with Bonferroni post hoc tests, and repeatability was assessed using intraclass correlation coefficients (ICC). A bootstrap method (5000 samples) evaluated the differences in repeatability. Agreement was analyzed using Bland-Altman quantile regression plots, and diagnostic performance was assessed via receiver operating characteristic (ROC) curves. MAIN OUTCOME MEASURES: Measurement accuracy, agreement, and repeatability. RESULTS: No significant differences in mean deviation (MD) and pattern standard deviation (PSD) were found between SS and SF; however, SQ demonstrated significantly better MD and PSD (p<0.001). Test duration was reduced by 57.8% for SF and 73.8% for SQ compared to SS (p<0.001). Bland-Altman analysis revealed that the mean differences in MD between SS and SF and between SS and SQ were -0.35 and -3.1, respectively. ICC values indicated excellent repeatability for SF, whereas SQ exhibited significantly lower ICCs (p<0.001). Among the 118 participants, 74 (62.7%) had visual field defects identified by SS, 72 (61.0%) by SF, and 31 (26.2%) by SQ. Sensitivity and specificity for SF were 83.8% and 77.3%, respectively, compared to 40.5% and 97.7% for SQ. ROC analysis yielded AUCs for MD of 0.89 for SF and 0.86 for SQ (p=0.306); for PSD, AUCs were 0.91 and 0.77 (p=0.005). The Conger's Kappa coefficient indicated fair agreement among the strategies. CONCLUSIONS: Both SF and SQ significantly shorten testing duration. SF yields results comparable to SS, while SQ may underestimate defects.

Authors

Institutions

Publication Details

Journal
PubMed
Published
2026-05-11
DOI
https://doi.org/10.1016/j.ogla.2026.05.001
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Comparative Evaluation of Two Rapid Strategies for Visual Field Testing: SITA Faster and SPARK Quick.

Chengtan Liu, X J Wang, Yize Chen, Xiaoqian Bian et al.
PubMed
Glaucoma and retinal disorders
article

A Comparative Evaluation of Two Rapid Strategies for Visual Field Testing: SITA Faster and SPARK Quick.

Chengtan Liu, X J Wang, Yize Chen, Xiaoqian Bian, Hao Lin, Yuanbo Liang, Cong Ye, Fan Lu
article en

Abstract

OBJECTIVES: This study compares the diagnostic performance of two rapid visual field testing strategies, SITA Faster (SF) and SPARK Quick (SQ), against the gold standard SITA Standard (SS). The measurement accuracy, agreement, and repeatability were evaluated. DESIGN: Prospective, cross-sectional study. PARTICIPANTS: A total of 118 eyes of 118 participants were enrolled. METHODS: Participants from a glaucoma clinic underwent multiple visual field tests. All subjects were naïve to SF and SQ, with SS 24-2 serving as the initial test. SF and SQ were then administered in random order. Among the 74 participants with visual field defects detected by SS, 49 underwent testing twice with both rapid strategies. Parameters were compared using one-way ANOVA with Bonferroni post hoc tests, and repeatability was assessed using intraclass correlation coefficients (ICC). A bootstrap method (5000 samples) evaluated the differences in repeatability. Agreement was analyzed using Bland-Altman quantile regression plots, and diagnostic performance was assessed via receiver operating characteristic (ROC) curves. MAIN OUTCOME MEASURES: Measurement accuracy, agreement, and repeatability. RESULTS: No significant differences in mean deviation (MD) and pattern standard deviation (PSD) were found between SS and SF; however, SQ demonstrated significantly better MD and PSD (p<0.001). Test duration was reduced by 57.8% for SF and 73.8% for SQ compared to SS (p<0.001). Bland-Altman analysis revealed that the mean differences in MD between SS and SF and between SS and SQ were -0.35 and -3.1, respectively. ICC values indicated excellent repeatability for SF, whereas SQ exhibited significantly lower ICCs (p<0.001). Among the 118 participants, 74 (62.7%) had visual field defects identified by SS, 72 (61.0%) by SF, and 31 (26.2%) by SQ. Sensitivity and specificity for SF were 83.8% and 77.3%, respectively, compared to 40.5% and 97.7% for SQ. ROC analysis yielded AUCs for MD of 0.89 for SF and 0.86 for SQ (p=0.306); for PSD, AUCs were 0.91 and 0.77 (p=0.005). The Conger's Kappa coefficient indicated fair agreement among the strategies. CONCLUSIONS: Both SF and SQ significantly shorten testing duration. SF yields results comparable to SS, while SQ may underestimate defects.

PubMed
Wenzhou Medical University (CN), Affiliated Eye Hospital of Wenzhou Medical College (CN)
Openalex Percentile: Top 4%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.