Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation

OBJECTIVE: The current method for mitral annuloplasty ring sizing has remained essentially unchanged for a half century and offers surgeons limited actionable information. A novel technique is illustrated and validated in a cadaver swine heart model. METHODS: Twenty cadaver swine hearts were used to validate the novel sizing device and method. We hypothesized that the paired-ring sizer would force annuloplasty sutures to conform to its exact shape and therefore accurately predict changes in coaptation length when used to "downsize" the mitral annulus. Anterior leaflet coaptation length was measured under conditions of left ventricular saline distention: (1) at baseline, (2) with the sizer in place temporarily holding the sutures in position, and (3) after final implantation of a (corresponding size) Physio II ring (Edwards Lifesciences, Irivne, CA, USA). The "predicted" (sizer) and "actual" (implanted Physio II ring) coaptation lengths were compared using Bland-Altman analysis to assess accuracy. RESULTS: A moderate (2.0 to 5.5 mm) increase in coaptation length was achieved in 19 hearts and no change (0.0 mm) in 1 heart. The initial predicted (sizer) coaptation lengths corresponded well (≤0.5 mm difference) with the final, actual (implanted ring) coaptation lengths across all 20 hearts. Bland-Altman plots are provided for (derived) coaptation length increases. The mean difference (95% confidence interval) indicated that predicted were slightly greater than actual (0.08 [-0.06, 0.21]) but not different than zero. Limits of agreement were from -0.50 to 0.65. There did not appear to be significant heteroscedasticity. CONCLUSIONS: The accuracy of a novel mitral annuloplasty ring-sizing device was validated using a cadaveric swine heart model. Coaptation lengths initially predicted by the sizer corresponded accurately with coaptation lengths observed upon implantation of a corresponding commercially available ring. Continued development of this technology should provide surgeons with precise control over the final coaptation length by previewing several different rings prior to committing to a final implant.

Authors

Institutions

Publication Details

Journal
Innovations Technology and Techniques in Cardiothoracic and Vascular Surgery
Published
2026-05-11
DOI
https://doi.org/10.1177/15569845261437227
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation

Nikola Dobrilovic
Innovations Technology and Techniques in Cardiothoracic and Vascular Surgery
Cardiac Valve Diseases and Treatments
article

Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation

Nikola Dobrilovic
article en

Abstract

OBJECTIVE: The current method for mitral annuloplasty ring sizing has remained essentially unchanged for a half century and offers surgeons limited actionable information. A novel technique is illustrated and validated in a cadaver swine heart model. METHODS: Twenty cadaver swine hearts were used to validate the novel sizing device and method. We hypothesized that the paired-ring sizer would force annuloplasty sutures to conform to its exact shape and therefore accurately predict changes in coaptation length when used to "downsize" the mitral annulus. Anterior leaflet coaptation length was measured under conditions of left ventricular saline distention: (1) at baseline, (2) with the sizer in place temporarily holding the sutures in position, and (3) after final implantation of a (corresponding size) Physio II ring (Edwards Lifesciences, Irivne, CA, USA). The "predicted" (sizer) and "actual" (implanted Physio II ring) coaptation lengths were compared using Bland-Altman analysis to assess accuracy. RESULTS: A moderate (2.0 to 5.5 mm) increase in coaptation length was achieved in 19 hearts and no change (0.0 mm) in 1 heart. The initial predicted (sizer) coaptation lengths corresponded well (≤0.5 mm difference) with the final, actual (implanted ring) coaptation lengths across all 20 hearts. Bland-Altman plots are provided for (derived) coaptation length increases. The mean difference (95% confidence interval) indicated that predicted were slightly greater than actual (0.08 [-0.06, 0.21]) but not different than zero. Limits of agreement were from -0.50 to 0.65. There did not appear to be significant heteroscedasticity. CONCLUSIONS: The accuracy of a novel mitral annuloplasty ring-sizing device was validated using a cadaveric swine heart model. Coaptation lengths initially predicted by the sizer corresponded accurately with coaptation lengths observed upon implantation of a corresponding commercially available ring. Continued development of this technology should provide surgeons with precise control over the final coaptation length by previewing several different rings prior to committing to a final implant.

Innovations Technology and Techniques in Cardiothoracic and Vascular Surgery
NorthShore University HealthSystem (US)
Responsible consumption and production
Openalex Percentile: Top 6%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.