Prompt gamma timing for range verification with carbon ion irradiation: first experimental measurements and comparison with Geant4 Monte Carlo simulations

Prompt Gamma Timing (PGT) is a promising technique for in vivo range verification in particle therapy, exploiting the time-of-flight between primary particles and prompt gamma rays emitted by nuclear interactions. PGT distribution is highly sensitive to beam energy and target density, which, under controlled detector positioning, enables real-time monitoring of particle range, detection of morphological changes, and support for adaptive treatment strategies.

Objective: This study investigates for the first time the application of PGT in carbon ion therapy. 

Approach: Measurements were performed using a dedicated detection system composed of a silicon strip sensor for primary ion timing and a LaBr3(Ce) scintillator read out by a SiPM matrix for secondary radiation. Carbon ion beams with energies of 166.41, 268.86, and 398.84 MeV/u irradiated a homogeneous 30.0 cm PMMA target at CNAO (Pavia, Italy). The secondary radiation detector was positioned at four off-beam positions to assess the robustness of the PGT technique. Monte Carlo simulations based on Geant4 were conducted for all configurations to evaluate agreement and predictive capability.

Main Results: A quantitative bin-by-bin comparison of experimental and simulated PGT intensities demonstrated strong agreement within the 95% confidence interval, with no incompatible bins at 166.41 MeV/u, at most 1% at 268.86 MeV/u, and up to 8% at 398.84 MeV/u, depending on detector position. Prompt gammas were identified as the dominant contribution to the detected signals, particularly for detector positions upstream with respect to the primary particle beam, minimizing signal contamination from neutrons and charged fragments. 

Significance: The validated experimental - simulation framework confirms the capability of the proposed PGT system to resolve energy-dependent differences and highlights its potential for detecting clinically relevant changes in the particle beam range, supporting further development toward real-time monitoring in carbon ion therapy.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-05-12
DOI
https://doi.org/10.1088/1361-6560/ae6cae
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Prompt gamma timing for range verification with carbon ion irradiation: first experimental measurements and comparison with Geant4 Monte Carlo simulations

S. Giordanengo, A. Vignati, M. Fadavi Mazinani, R. Cirio et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Prompt gamma timing for range verification with carbon ion irradiation: first experimental measurements and comparison with Geant4 Monte Carlo simulations

S. Giordanengo, A. Vignati, M. Fadavi Mazinani, R. Cirio, R. Sacchi, Félix Más Milián, Miguel David Fernandez Moreira, Francesco Pennazio, Simone Savazzi, Emanuele Maria Data, Marco Pullia, Veronica Ferrero, Sahar Ranjbar, Iram B Rivas Ortiz, Elisa Fiorina, Piergiorgio Cerello, Diango Montalvan Olivares
article en

Abstract

Prompt Gamma Timing (PGT) is a promising technique for in vivo range verification in particle therapy, exploiting the time-of-flight between primary particles and prompt gamma rays emitted by nuclear interactions. PGT distribution is highly sensitive to beam energy and target density, which, under controlled detector positioning, enables real-time monitoring of particle range, detection of morphological changes, and support for adaptive treatment strategies.

Objective: This study investigates for the first time the application of PGT in carbon ion therapy. 

Approach: Measurements were performed using a dedicated detection system composed of a silicon strip sensor for primary ion timing and a LaBr3(Ce) scintillator read out by a SiPM matrix for secondary radiation. Carbon ion beams with energies of 166.41, 268.86, and 398.84 MeV/u irradiated a homogeneous 30.0 cm PMMA target at CNAO (Pavia, Italy). The secondary radiation detector was positioned at four off-beam positions to assess the robustness of the PGT technique. Monte Carlo simulations based on Geant4 were conducted for all configurations to evaluate agreement and predictive capability.

Main Results: A quantitative bin-by-bin comparison of experimental and simulated PGT intensities demonstrated strong agreement within the 95% confidence interval, with no incompatible bins at 166.41 MeV/u, at most 1% at 268.86 MeV/u, and up to 8% at 398.84 MeV/u, depending on detector position. Prompt gammas were identified as the dominant contribution to the detected signals, particularly for detector positions upstream with respect to the primary particle beam, minimizing signal contamination from neutrons and charged fragments. 

Significance: The validated experimental - simulation framework confirms the capability of the proposed PGT system to resolve energy-dependent differences and highlights its potential for detecting clinically relevant changes in the particle beam range, supporting further development toward real-time monitoring in carbon ion therapy.

Physics in Medicine and Biology
Istituto Nazionale di Fisica Nucleare, Sezione di Torino (IT), National Center for Oncological Hadrontherapy (IT), University of Turin (IT)
Instituto Nazionale di Fisica Nucleare, Commissione Scientifica Nazionale 5, Instituto Nazionale di Fisica Nucleare
Openalex Percentile: Top 6%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.