Lot-to-lot variability of a high-sensitivity cardiac troponin I assay: clinical implications

OBJECTIVES: High-sensitivity cardiac troponin I (hs-cTnI) assays are pivotal for diagnosing acute coronary syndromes and implementing accelerated rule-in/rule-out algorithms in the emergency department (ED). This study assessed the analytical and clinical performance of the Mindray hs-cTnI chemiluminescent immunoassay in comparison with the assays routinely adopted at the University Hospital of Padova (AOUP; Access, Beckman Coulter) and at Bassano Hospital (ULSS7; Architect, Abbott). Particular attention was given to reagent and calibrator lot-to-lot variability and its potential clinical impact. METHODS: The study included 400 ED patients with suspected myocardial injury (213 at AOUP [P1] and 187 at ULSS7 [P2]), yielding a total of 823 lithium-heparin plasma samples collected at baseline, 2 h, and, when available, at a third time point. Lot-to-lot comparability was examined in 424 samples using two reagent/calibrator lot combinations (R1/C1 and R2/C2) and further explored in 51 additional samples tested with four lot combinations. RESULTS: Within-laboratory imprecision was 2.03 % at 2.45 ng/L and 2.25 % at 4.76 ng/L. Among patients discharged from ED, the proportion of baseline values exceeding the 99th percentile was lower with the Mindray assay (P1: 16.2 %, P2: 15.1 %) compared with routine methods (P1: 26.8 %, P2: 17.8 %). Furthermore, diagnostic agreement was higher with Architect than with Access (Cohen's kappa 0.93 vs. 0.77). Lot-to-lot comparison demonstrated excellent concordance between R1/C1 and R2/C2 (bias 0.73 %; kappa 0.98). CONCLUSIONS: The evaluated assay demonstrated robust analytical and clinical performance, together with excellent lot-to-lot stability. Nevertheless, ongoing surveillance of reagent and calibrator lot changes remains advisable to ensure sustained analytical consistency and clinical reliability.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-05-12
DOI
https://doi.org/10.1515/cclm-2026-0315
Primary Topic
Acute Myocardial Infarction Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Lot-to-lot variability of a high-sensitivity cardiac troponin I assay: clinical implications

Mariela Marinova, Luca Licchelli, Chiara Cosma, M Plebani et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Acute Myocardial Infarction Research
article

Lot-to-lot variability of a high-sensitivity cardiac troponin I assay: clinical implications

Mariela Marinova, Luca Licchelli, Chiara Cosma, M Plebani, Elisa Pangrazzi, M M Mion, Ilaria Talli, Francesca Merola, Martina Zaninotto
article en

Abstract

OBJECTIVES: High-sensitivity cardiac troponin I (hs-cTnI) assays are pivotal for diagnosing acute coronary syndromes and implementing accelerated rule-in/rule-out algorithms in the emergency department (ED). This study assessed the analytical and clinical performance of the Mindray hs-cTnI chemiluminescent immunoassay in comparison with the assays routinely adopted at the University Hospital of Padova (AOUP; Access, Beckman Coulter) and at Bassano Hospital (ULSS7; Architect, Abbott). Particular attention was given to reagent and calibrator lot-to-lot variability and its potential clinical impact. METHODS: The study included 400 ED patients with suspected myocardial injury (213 at AOUP [P1] and 187 at ULSS7 [P2]), yielding a total of 823 lithium-heparin plasma samples collected at baseline, 2 h, and, when available, at a third time point. Lot-to-lot comparability was examined in 424 samples using two reagent/calibrator lot combinations (R1/C1 and R2/C2) and further explored in 51 additional samples tested with four lot combinations. RESULTS: Within-laboratory imprecision was 2.03 % at 2.45 ng/L and 2.25 % at 4.76 ng/L. Among patients discharged from ED, the proportion of baseline values exceeding the 99th percentile was lower with the Mindray assay (P1: 16.2 %, P2: 15.1 %) compared with routine methods (P1: 26.8 %, P2: 17.8 %). Furthermore, diagnostic agreement was higher with Architect than with Access (Cohen's kappa 0.93 vs. 0.77). Lot-to-lot comparison demonstrated excellent concordance between R1/C1 and R2/C2 (bias 0.73 %; kappa 0.98). CONCLUSIONS: The evaluated assay demonstrated robust analytical and clinical performance, together with excellent lot-to-lot stability. Nevertheless, ongoing surveillance of reagent and calibrator lot changes remains advisable to ensure sustained analytical consistency and clinical reliability.

Clinical Chemistry and Laboratory Medicine (CCLM)
University of Padua (IT), Ospedale San Bassiano (IT)
Good health and well-being
Openalex Percentile: Top 6%
Acute Myocardial Infarction Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.