Relative Motion of Cementless Tibial Trays: Time-Elapsed Micro-CT and DVC Analysis During Stair Descent and Deep-Knee-Bend

An interference fit is employed for cementless tibial total knee arthroplasty components to ensure stability post-operatively. Relative motion between implant and underlying bone should be minimised, facilitating osseointegration. Previous experimental limitations have prevented quantification of this motion across the entire bone-implant interface. Furthermore, the intended (nominal) interference fit differs to that actually achieved. This study aimed to 1) experimentally quantify the relative motion across the entire interface of a commercially available cementless tibial component and underlying bone, for cadaveric tibiae through micro-computed tomography (micro-CT) imaging and digital volume correlation (DVC); and 2) assess the relationship between relative motion, actual interference fit, post-impaction strain and bone volume fraction (BV/TV).Seven cadaveric tibiae were micro-CT scanned when intact, following resection and, once implanted with a cementless titanium tibial component (Attune, DePuy Synthes), during two time-elapsed mechanical loading sequences that replicated stair descent (SD, 0.0 - 2.5 bodyweight) and deep knee bend (DKB, 0.0-3.5 bodyweight). Actual interference fit was quantified from micro-CT datasets. DVC was employed to extract the post-impaction strain field and, for each load case, the relative motion at the bone-implant interface.A sagittal rocking motion of the tray was detected in both SD and DKB, that increased with increased load applied. Tray lift-off was enhanced in the central- and lateral-anterior regions during DKB. Reduced relative motion was significantly related to higher interference in most cases, whilst it was not related to post-impaction strain or BV/TV. This study provides a comprehensive insight into the mechanical environment surrounding a cementless tibial tray in the immediate post-implantation period.

Authors

Institutions

Publication Details

Journal
Annals of Biomedical Engineering
Published
2026-05-12
DOI
https://doi.org/10.1007/s10439-026-04102-z
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Relative Motion of Cementless Tibial Trays: Time-Elapsed Micro-CT and DVC Analysis During Stair Descent and Deep-Knee-Bend

M. Taylor, M. Awadalla, S. Rapagna, L. S. Wearne et al.
Annals of Biomedical Engineering
Total Knee Arthroplasty Outcomes
article

Relative Motion of Cementless Tibial Trays: Time-Elapsed Micro-CT and DVC Analysis During Stair Descent and Deep-Knee-Bend

M. Taylor, M. Awadalla, S. Rapagna, L. S. Wearne, G. Keene, E. Perilli
article en

Abstract

An interference fit is employed for cementless tibial total knee arthroplasty components to ensure stability post-operatively. Relative motion between implant and underlying bone should be minimised, facilitating osseointegration. Previous experimental limitations have prevented quantification of this motion across the entire bone-implant interface. Furthermore, the intended (nominal) interference fit differs to that actually achieved. This study aimed to 1) experimentally quantify the relative motion across the entire interface of a commercially available cementless tibial component and underlying bone, for cadaveric tibiae through micro-computed tomography (micro-CT) imaging and digital volume correlation (DVC); and 2) assess the relationship between relative motion, actual interference fit, post-impaction strain and bone volume fraction (BV/TV).Seven cadaveric tibiae were micro-CT scanned when intact, following resection and, once implanted with a cementless titanium tibial component (Attune, DePuy Synthes), during two time-elapsed mechanical loading sequences that replicated stair descent (SD, 0.0 - 2.5 bodyweight) and deep knee bend (DKB, 0.0-3.5 bodyweight). Actual interference fit was quantified from micro-CT datasets. DVC was employed to extract the post-impaction strain field and, for each load case, the relative motion at the bone-implant interface.A sagittal rocking motion of the tray was detected in both SD and DKB, that increased with increased load applied. Tray lift-off was enhanced in the central- and lateral-anterior regions during DKB. Reduced relative motion was significantly related to higher interference in most cases, whilst it was not related to post-impaction strain or BV/TV. This study provides a comprehensive insight into the mechanical environment surrounding a cementless tibial tray in the immediate post-implantation period.

Annals of Biomedical Engineering
Flinders University (AU), Sports Medicine Australia (AU)
Australian National Fabrication Facility, Flinders University, Australian Research Council
Openalex Percentile: Top 5%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.