ThermiQuantTM MegaScan: high-throughput isothermal reactor with quantitative colorimetric readout for paper-based nucleic acid amplification tests

Isothermal nucleic acid amplification tests (NAATs), such as loop-mediated isothermal amplification (LAMP) implemented on microfluidic paper-based analytical devices (µPADs), enable inexpensive and rapid (≤60 min) colorimetric molecular diagnostics; however, no existing instrument supports high-throughput (>100 reactions) quantitative analysis of colorimetric isothermal assays on paper substrates under controlled laboratory conditions. To address this gap, we developed ThermiQuant™ MegaScan, a scanner- and water-bath–based platform that accommodates a 160-reaction µPAD cartridge, maintains uniform incubation at 65 ± 0.5 °C, and enables real-time imaging every 30 s. We also developed accompanying software, Amplimetrics™, for automated µPAD detection and kinetic colorimetric analysis. Using paper-based colorimetric LAMP targeting the SARS-CoV-2 orf7ab region, the assay achieved a limit of detection at 95% probability (LOD95) of 34 copies per reaction (5 copies/µL) and a limit of quantification (LOQ) of 250 copies per reaction (33 copies/µL) using purified synthetic DNA targets, and achieved 72% sensitivity and 100% specificity relative to digital PCR (dPCR) for diluted human nasopharyngeal (NP) swab virus samples. We further evaluated the effects of viral and universal transport media (VTM/UTM) on assay performance and found that linear calibration derived from synthetic targets do not reliably translate to clinical samples in these media. Together, these results establish ThermiQuant™ MegaScan as a high-throughput laboratory research platform for standardized evaluation, optimization, and benchmarking of paper-based colorimetric nucleic acid amplification assays.

Authors

Institutions

Publication Details

Journal
npj Biosensing
Published
2026-05-12
DOI
https://doi.org/10.1038/s44328-026-00099-6
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

ThermiQuantTM MegaScan: high-throughput isothermal reactor with quantitative colorimetric readout for paper-based nucleic acid amplification tests

Bibek Raut, Josiah Levi Davidson, Mohit S. Verma, Ryan F. Relich et al.
npj Biosensing
Biosensors and Analytical Detection
article

ThermiQuantTM MegaScan: high-throughput isothermal reactor with quantitative colorimetric readout for paper-based nucleic acid amplification tests

Bibek Raut, Josiah Levi Davidson, Mohit S. Verma, Ryan F. Relich, J. Pasternak, Gopal Palla, Andrew Fleck, Virendra Kumar, Jon P. Schoonmaker, Bilal Ahmed
article en

Abstract

Isothermal nucleic acid amplification tests (NAATs), such as loop-mediated isothermal amplification (LAMP) implemented on microfluidic paper-based analytical devices (µPADs), enable inexpensive and rapid (≤60 min) colorimetric molecular diagnostics; however, no existing instrument supports high-throughput (>100 reactions) quantitative analysis of colorimetric isothermal assays on paper substrates under controlled laboratory conditions. To address this gap, we developed ThermiQuant™ MegaScan, a scanner- and water-bath–based platform that accommodates a 160-reaction µPAD cartridge, maintains uniform incubation at 65 ± 0.5 °C, and enables real-time imaging every 30 s. We also developed accompanying software, Amplimetrics™, for automated µPAD detection and kinetic colorimetric analysis. Using paper-based colorimetric LAMP targeting the SARS-CoV-2 orf7ab region, the assay achieved a limit of detection at 95% probability (LOD95) of 34 copies per reaction (5 copies/µL) and a limit of quantification (LOQ) of 250 copies per reaction (33 copies/µL) using purified synthetic DNA targets, and achieved 72% sensitivity and 100% specificity relative to digital PCR (dPCR) for diluted human nasopharyngeal (NP) swab virus samples. We further evaluated the effects of viral and universal transport media (VTM/UTM) on assay performance and found that linear calibration derived from synthetic targets do not reliably translate to clinical samples in these media. Together, these results establish ThermiQuant™ MegaScan as a high-throughput laboratory research platform for standardized evaluation, optimization, and benchmarking of paper-based colorimetric nucleic acid amplification assays.

npj Biosensing
Indiana University Health (US), University of Kentucky (US), Purdue University West Lafayette (US), Indiana University School of Medicine, Indiana University – Purdue University Indianapolis (US)
Foundation for Food and Agriculture Research, National Institute of Food and Agriculture, Animal and Plant Health Inspection Service, School of Medicine, Indiana University, College of Agriculture, Purdue University, Wabash Heartland Innovation Network
Clean water and sanitation
Openalex Percentile: Top 9%
Biosensors and Analytical Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.