Variation in Preoperative Refraction Results Among Refractive Surgery Patients: A Comparison of Five Measurement Methods

Background: Accurate preoperative refraction is critical in refractive surgery planning. Manifest refraction (MR) remains the clinical gold standard but is time-consuming and operator-dependent. Alternative objective methods, including wavefront aberrometry and automated refraction, offer speed and objectivity, yet their agreement with MR varies, particularly regarding astigmatic components. Identifying which alternatives most closely approximate MR has direct clinical implications. Objective: To compare preoperative spherical equivalent (SE) refraction obtained by five measurement methods and identify which most closely agree with MR as potential clinical alternatives. Methods: Retrospective comparative study of 104 eyes (52 patients) at IIUM Eye Specialist Clinic (January-May 2021). Five methods were compared: MR, Wavefront Supported Custom Ablation (WASCA) aberrometer, Zeiss i.Profiler (ZIP), Righton Retinomax Handheld (RRH) autorefractor, and focimeter (FOCI). Pairwise comparisons used paired t-tests on SE values. Results: Significant pairwise differences were found between all five methods (P < 0.05), except ZIP vs. RRH (mean difference (MD) = 0.03 diopters (D); P = 0.753). The two methods with the smallest mean differences from MR were ZIP (MD = 0.20 D) and RRH (MD = 0.22 D), both below the 0.25 D clinical significance threshold. The largest discrepancy was between WASCA and FOCI (MD = 1.44 D). Conclusion: Both the ZIP and RRH autorefractors produced SE values in close agreement with MR. ZIP is preferred due to superior positional control, yielding more reproducible astigmatic readings. Routine autorefraction with ZIP may supplement MR in preoperative refractive surgery assessment.

Authors

Publication Details

Journal
Cureus
Published
2026-05-11
DOI
https://doi.org/10.7759/cureus.108693
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Variation in Preoperative Refraction Results Among Refractive Surgery Patients: A Comparison of Five Measurement Methods

Azuwan Musa, Abd Bari Muhd-Syafi, Muhammad Yusuf Abdurrahman, Khairidzan Mohd. Kamal
Cureus
Ophthalmology and Visual Impairment Studies
article

Variation in Preoperative Refraction Results Among Refractive Surgery Patients: A Comparison of Five Measurement Methods

Azuwan Musa, Abd Bari Muhd-Syafi, Muhammad Yusuf Abdurrahman, Khairidzan Mohd. Kamal
article en

Abstract

Background: Accurate preoperative refraction is critical in refractive surgery planning. Manifest refraction (MR) remains the clinical gold standard but is time-consuming and operator-dependent. Alternative objective methods, including wavefront aberrometry and automated refraction, offer speed and objectivity, yet their agreement with MR varies, particularly regarding astigmatic components. Identifying which alternatives most closely approximate MR has direct clinical implications. Objective: To compare preoperative spherical equivalent (SE) refraction obtained by five measurement methods and identify which most closely agree with MR as potential clinical alternatives. Methods: Retrospective comparative study of 104 eyes (52 patients) at IIUM Eye Specialist Clinic (January-May 2021). Five methods were compared: MR, Wavefront Supported Custom Ablation (WASCA) aberrometer, Zeiss i.Profiler (ZIP), Righton Retinomax Handheld (RRH) autorefractor, and focimeter (FOCI). Pairwise comparisons used paired t-tests on SE values. Results: Significant pairwise differences were found between all five methods (P < 0.05), except ZIP vs. RRH (mean difference (MD) = 0.03 diopters (D); P = 0.753). The two methods with the smallest mean differences from MR were ZIP (MD = 0.20 D) and RRH (MD = 0.22 D), both below the 0.25 D clinical significance threshold. The largest discrepancy was between WASCA and FOCI (MD = 1.44 D). Conclusion: Both the ZIP and RRH autorefractors produced SE values in close agreement with MR. ZIP is preferred due to superior positional control, yielding more reproducible astigmatic readings. Routine autorefraction with ZIP may supplement MR in preoperative refractive surgery assessment.

Cureus
Quality Education
Openalex Percentile: Top 6%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.