Comparative Analysis of the LOBO Device and Coils for Proximal Splenic Artery Embolization: A Multi-Institutional Retrospective Study

PURPOSE: To evaluate the technical and clinical efficacy of the low-profile braided occluder (LOBO) device for proximal splenic artery embolization (SAE) and compare its performance with conventional coil embolization. MATERIALS AND METHODS: This retrospective multi-institutional study included 44 consecutive patients who underwent proximal SAE between June 2023 and October 2025. Demographic data, etiology, embolic device type, fluoroscopy time, radiation dose, and outcomes were recorded. Technical success was defined as complete angiographic occlusion using the index device. Clinical success was defined as the absence of rebleeding or reintervention within 30 days. RESULTS: Of the 44 patients (mean age, 43.2 years; 39% females), 22 were treated using the LOBO device and 22 with coils. Trauma was the most common etiology (n = 37, 84.1%). There were no clinical or demographic characteristic differences between the LOBO and coil groups. Proximal SAE was achieved with a single LOBO device in all patients. The mean number of coils used was 4.6 per case (median 3.5), p < 0.001. Median fluoroscopy time was significantly reduced with LOBO compared with coils (11.65 vs 18.9 min; p = 0.01). Radiation dose was significantly lower as well in the LOBO group compared with the coil group (128.4 vs 198.5 mGy; p = 0.04). Clinical success was not significantly different between the groups, with the LOBO group achieving 100% and the coil group 90.9% (p > 0.05). Splenic infarction was significantly higher with coils than with LOBO (33.3% vs 22.7%; p = 0.046). No major adverse events occurred, and minor events were self-limited. One death in each group was attributable to polytrauma and unrelated to the procedure. CONCLUSION: The LOBO device provides safe, effective proximal SAE with significantly reduced fluoroscopy time and radiation dose compared with coils. Larger prospective studies are warranted. LEVEL OF EVIDENCE: Level 4, study (retrospective, standard quality).

Authors

Institutions

Publication Details

Journal
CardioVascular and Interventional Radiology
Published
2026-05-12
DOI
https://doi.org/10.1007/s00270-026-04460-0
Primary Topic
Abdominal vascular conditions and treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Analysis of the LOBO Device and Coils for Proximal Splenic Artery Embolization: A Multi-Institutional Retrospective Study

Massoud Allahyari, Roshan Valentine, Bahareh Gholami, Alexander Villalobos et al.
CardioVascular and Interventional Radiology
Abdominal vascular conditions and treatments
article

Comparative Analysis of the LOBO Device and Coils for Proximal Splenic Artery Embolization: A Multi-Institutional Retrospective Study

Massoud Allahyari, Roshan Valentine, Bahareh Gholami, Alexander Villalobos, Chaitanya Ahuja, Harrison Bieber, Gloria Salazar, Johannes L. Du Pisane, Ali Afrasiabi, Nicholas Thomas, Sandra Gad, David Mauro, Nima Kokabi
article en

Abstract

PURPOSE: To evaluate the technical and clinical efficacy of the low-profile braided occluder (LOBO) device for proximal splenic artery embolization (SAE) and compare its performance with conventional coil embolization. MATERIALS AND METHODS: This retrospective multi-institutional study included 44 consecutive patients who underwent proximal SAE between June 2023 and October 2025. Demographic data, etiology, embolic device type, fluoroscopy time, radiation dose, and outcomes were recorded. Technical success was defined as complete angiographic occlusion using the index device. Clinical success was defined as the absence of rebleeding or reintervention within 30 days. RESULTS: Of the 44 patients (mean age, 43.2 years; 39% females), 22 were treated using the LOBO device and 22 with coils. Trauma was the most common etiology (n = 37, 84.1%). There were no clinical or demographic characteristic differences between the LOBO and coil groups. Proximal SAE was achieved with a single LOBO device in all patients. The mean number of coils used was 4.6 per case (median 3.5), p < 0.001. Median fluoroscopy time was significantly reduced with LOBO compared with coils (11.65 vs 18.9 min; p = 0.01). Radiation dose was significantly lower as well in the LOBO group compared with the coil group (128.4 vs 198.5 mGy; p = 0.04). Clinical success was not significantly different between the groups, with the LOBO group achieving 100% and the coil group 90.9% (p > 0.05). Splenic infarction was significantly higher with coils than with LOBO (33.3% vs 22.7%; p = 0.046). No major adverse events occurred, and minor events were self-limited. One death in each group was attributable to polytrauma and unrelated to the procedure. CONCLUSION: The LOBO device provides safe, effective proximal SAE with significantly reduced fluoroscopy time and radiation dose compared with coils. Larger prospective studies are warranted. LEVEL OF EVIDENCE: Level 4, study (retrospective, standard quality).

CardioVascular and Interventional Radiology
University of North Carolina at Chapel Hill (US), University of North Carolina Health Care (US), St. George's University (GD), Louisiana State University Health Sciences Center Shreveport (US)
Good health and well-being
Openalex Percentile: Top 5%
Abdominal vascular conditions and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.