Hemodynamic performance of valve-in-valve-in-valve implantation for transcatheter heart valve: an in vitro simulation study
BACKGROUND&OBJECTIVE: Valve-in-valve-in-valve (ViViV) implantation will be considered after the second valve deterioration as an increasing number of patients aged < 60 years undergo transcatheter aortic valve replacement (TAVR). We aimed to investigate the hemodynamic outcomes after ViViV procedures. METHOD: Evolut and Venus A with size of 23 mm and 26 mm were selected for the testing. Cardiovascular hydrodynamic tests were conducted using the in Vitro Pulse Duplicator system. RESULT: ViViV configuration of 26 mm-in-26 mm-in-26 mm demonstrates a larger effective orifice area (EOA) than the 23 mm-in-26 mm-in-26 mm configuration. The transvalvular mean pressure gradient (MPG) of three 26 mm combination were 8.3 ± 1.9 mmHg for Venus-A and 13.2 ± 1.6 mmHg for Evolut, which were acceptable. CONCLUSION: The ViViV valve combination employing three 26 mm Venus-A or Evolut self-expanding valves, as opposed to the 23 mm-in-26 mm-in-26 mm configuration, exhibited favorable hemodynamic performance as demonstrated by in the vitro testing.
Authors
- Wentao Yan (ORCID: https://orcid.org/0000-0001-8480-6375)
- Daxin Zhou
- Dawei Lin
- Junbo Ge
- Jian Guo
- Yuliang Long
- Wenzhi Pan
Institutions
- Sun Yat-sen University (CN)
- Chinese Academy of Medical Sciences & Peking Union Medical College (CN)
- Fudan University (CN)
- Zhongshan Hospital (CN)
- The First Affiliated Hospital, Sun Yat-sen University (CN)
- Shanghai Clinical Research Center (CN)
Publication Details
- Journal
- BMC Cardiovascular Disorders
- Published
- 2026-05-12
- DOI
- https://doi.org/10.1186/s12872-026-05867-w
- Primary Topic
- Cardiac Valve Diseases and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Natural Science Foundation of China