Objectives: The aim of this study was to compare a widely applied table-top digital non-mydriatic camera (Topcon TRC-NW-8) with a handheld digital non-mydriatic camera (Optomed Aurora IQ) regarding the quantitative assessment of the retinal microcirculation using established biomarkers: central retinal arteriolar equivalent (CRAE), central retinal venular equivalent (CRVE) and arterio-venous ratio (AVR). Methods: The present cross-sectional study included 26 randomly selected participants (51 eyes) who underwent retinal imaging of both eyes with the two devices and were analyzed using a static retinal vessel analyzer. Results: The mean differences in CRAE, CRVE and AVR between the two devices (Topcon/Aurora) were 24.96 ± 11.7, 22.7 ± 11.7 and 0.026 ± 0.045, respectively. Strong correlations were observed between devices (r = 0.84 for CRAE, 0.75 for CRVE and 0.83 for AVR; all p < 0.001), with high agreement as indicated by ICC values (0.91, 0.85, and 0.90, respectively). Bland–Altman plots indicated evidence of systemic bias (95% within 2 SD) with no proportional bias, as the differences were consistently distributed across the range of average values. Regression-based equations were derived to approximate the transformation of measurements between devices. Conclusions: The handheld fundus camera demonstrates strong correlation and good relative agreement with the table-top device; however, a consistent device-dependent bias limits the direct interchangeability of absolute measurements. The derived transformation equations may facilitate approximate cross-device comparison, although external validation is required. These findings support the complementary use of handheld devices and highlight the need for calibration strategies when integrating measurements across platforms.

Authors

Institutions

Publication Details

Journal
Reports — Medical Cases Images and Videos
Published
2026-05-11
DOI
https://doi.org/10.3390/reports9020147
Primary Topic
Retinal Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative Comparison of a Handheld and a Table-Top Fundus Camera for Retinal Microvascular Assessment

Petros Paul Sfikakis, Christina Konstantaki, Lazaros K. Yofoglu, A. Argyris et al.
Reports — Medical Cases Images and Videos
Retinal Imaging and Analysis
article

Quantitative Comparison of a Handheld and a Table-Top Fundus Camera for Retinal Microvascular Assessment

Petros Paul Sfikakis, Christina Konstantaki, Lazaros K. Yofoglu, A. Argyris, Evaggelia K. Aissopou, Chrysoula Moustou, K S Stamatelopoulos, Athanase D. Protogerou, Irini Chatziralli, Georgios Zervas
article en

Abstract

Objectives: The aim of this study was to compare a widely applied table-top digital non-mydriatic camera (Topcon TRC-NW-8) with a handheld digital non-mydriatic camera (Optomed Aurora IQ) regarding the quantitative assessment of the retinal microcirculation using established biomarkers: central retinal arteriolar equivalent (CRAE), central retinal venular equivalent (CRVE) and arterio-venous ratio (AVR). Methods: The present cross-sectional study included 26 randomly selected participants (51 eyes) who underwent retinal imaging of both eyes with the two devices and were analyzed using a static retinal vessel analyzer. Results: The mean differences in CRAE, CRVE and AVR between the two devices (Topcon/Aurora) were 24.96 ± 11.7, 22.7 ± 11.7 and 0.026 ± 0.045, respectively. Strong correlations were observed between devices (r = 0.84 for CRAE, 0.75 for CRVE and 0.83 for AVR; all p < 0.001), with high agreement as indicated by ICC values (0.91, 0.85, and 0.90, respectively). Bland–Altman plots indicated evidence of systemic bias (95% within 2 SD) with no proportional bias, as the differences were consistently distributed across the range of average values. Regression-based equations were derived to approximate the transformation of measurements between devices. Conclusions: The handheld fundus camera demonstrates strong correlation and good relative agreement with the table-top device; however, a consistent device-dependent bias limits the direct interchangeability of absolute measurements. The derived transformation equations may facilitate approximate cross-device comparison, although external validation is required. These findings support the complementary use of handheld devices and highlight the need for calibration strategies when integrating measurements across platforms.

Reports — Medical Cases Images and VideosVol. 9(2)
National and Kapodistrian University of Athens (GR), Alexandra Hospital (GR), Laiko General Hospital of Athens (GR), Hellenic Health Foundation (GR)
Good health and well-being
Openalex Percentile: Top 6%
Retinal Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.