Systematic Benchmarking of Spectral Demodulation Methods for Ball Resonator and Hybrid FPI–Ball Resonator Sensors for Multiparameter Physiological Monitoring
Ball resonator optical fiber sensors (OFSs) can offer multiparameter sensing capability, but their non-periodic and low-finesse reflection spectra make conventional spectral demodulation unreliable. This work proposes two sensor configurations: (i) a ball resonator and (ii) a hybrid sensor integrating a Fabry–Pérot interferometer (FPI) with a ball resonator, and compares their performance for multiparameter physiological monitoring using the Karhunen–Loève transform (KLT). The sensors were evaluated for glucose concentration (0–3 mg/mL), temperature (20–55 °C), and pH (3–9) monitoring. The ball resonator sensor, paired with KLT, achieved high linearity across all measurands (R2 = 0.989, 0.919, and 0.838 in response to glucose, temperature, and pH, respectively). The hybrid sensor exhibited a higher glucose sensitivity (6.15 a.u./(mg/mL)) compared to the ball resonator (3.77 a.u./(mg/mL)), resulting in limits of detection (LODs) of 2.53 mM and 4.19 mM, respectively. In contrast, the ball resonator sensor demonstrated better sensitivity for temperature and pH sensing. Furthermore, we present a comprehensive benchmarking framework of seven spectral demodulation methods for OFSs. The results demonstrated that KLT consistently provides robust demodulation performance and highlighted the potential of KLT for multiparameter physiological sensing applications.
Authors
- Natsnet Bereket Tecle
- M. Fatima Domingues
Institutions
- Khalifa University of Science and Technology (AE)
- Instituto de Telecomunicações (PT)
- University of Aveiro (PT)
Publication Details
- Journal
- Biosensors
- Published
- 2026-05-11
- DOI
- https://doi.org/10.3390/bios16050278
- Primary Topic
- Advanced Fiber Optic Sensors
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Khalifa University of Science, Technology and Research