Noise-Matched Blending Level Selection for 1024-Matrix CT Images Using Hybrid-Iterative Reconstruction: Comparison With 512-Matrix Images

OBJECTIVE: To identify the noise-matched hybrid-iterative reconstruction (IR) blending level for 1024-matrix images with noise equivalent to that of 512-matrix images while preserving comparable image quality. METHODS: In the phantom study, we identified the noise-matched hybrid-IR blending level for 1024-matrix images that achieved a noise level equivalent to 512-matrix images reconstructed with adaptive statistical iterative reconstruction-Veo (ASiR-V) 30%. This retrospective study included 63 patients (40 women; median age: 74y) who had undergone single-energy pancreatic protocol computed tomography (CT) from August 2023 to June 2024. For each patient, 3 image sets at the pancreatic phase were reconstructed: 512-matrix with ASiR-V 30% (group 1), 1024-matrix with ASiR-V 30% (group 2), and 1024-matrix with the noise-matched ASiR-V blending level (group 3). Background noise was measured as the SD of CT attenuation in the abdominal subcutaneous fat. Edge-rise slope (ERS) at the pancreas-fat interface was used to quantify edge sharpness. Two radiologists assessed overall image quality using a 5-point scale. RESULTS: The phantom study identified 60% as the noise-matched blending level for group 3. Background noise was significantly lower in the order of groups 3, 1, and 2, respectively (P < 0.001). ERS value was significantly higher in the order of groups 2, 3, and 1, respectively (P < 0.001). Overall image quality was significantly superior in the order of groups 3, 1, and 2, respectively (P < 0.001 for both radiologists). CONCLUSIONS: Increasing the matrix size leads to greater image noise with conventional reconstruction; however, applying a noise-matched hybrid-IR blending level allowed 1024-matrix images to maintain image quality while improving edge sharpness.

Authors

Institutions

Publication Details

Journal
Journal of Computer Assisted Tomography
Published
2026-05-11
DOI
https://doi.org/10.1097/rct.0000000000001878
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Noise-Matched Blending Level Selection for 1024-Matrix CT Images Using Hybrid-Iterative Reconstruction: Comparison With 512-Matrix Images

Tetsuro Kaga, Shingo Omata, Yukako Iritani, Hirohiko Imai et al.
Journal of Computer Assisted Tomography
Advanced X-ray and CT Imaging
article

Noise-Matched Blending Level Selection for 1024-Matrix CT Images Using Hybrid-Iterative Reconstruction: Comparison With 512-Matrix Images

Tetsuro Kaga, Shingo Omata, Yukako Iritani, Hirohiko Imai, Yukiko Takai, 三好利治, Masashi Asano, Abdelazim E. Elhelaly, Yoshifumi Noda, Masayuki Matsuo, Takeshi Iwata, Naoto Ozawa, Akio Ito, Hiroki Kato, Nobuyuki Kawai
article en

Abstract

OBJECTIVE: To identify the noise-matched hybrid-iterative reconstruction (IR) blending level for 1024-matrix images with noise equivalent to that of 512-matrix images while preserving comparable image quality. METHODS: In the phantom study, we identified the noise-matched hybrid-IR blending level for 1024-matrix images that achieved a noise level equivalent to 512-matrix images reconstructed with adaptive statistical iterative reconstruction-Veo (ASiR-V) 30%. This retrospective study included 63 patients (40 women; median age: 74y) who had undergone single-energy pancreatic protocol computed tomography (CT) from August 2023 to June 2024. For each patient, 3 image sets at the pancreatic phase were reconstructed: 512-matrix with ASiR-V 30% (group 1), 1024-matrix with ASiR-V 30% (group 2), and 1024-matrix with the noise-matched ASiR-V blending level (group 3). Background noise was measured as the SD of CT attenuation in the abdominal subcutaneous fat. Edge-rise slope (ERS) at the pancreas-fat interface was used to quantify edge sharpness. Two radiologists assessed overall image quality using a 5-point scale. RESULTS: The phantom study identified 60% as the noise-matched blending level for group 3. Background noise was significantly lower in the order of groups 3, 1, and 2, respectively (P < 0.001). ERS value was significantly higher in the order of groups 2, 3, and 1, respectively (P < 0.001). Overall image quality was significantly superior in the order of groups 3, 1, and 2, respectively (P < 0.001 for both radiologists). CONCLUSIONS: Increasing the matrix size leads to greater image noise with conventional reconstruction; however, applying a noise-matched hybrid-IR blending level allowed 1024-matrix images to maintain image quality while improving edge sharpness.

Journal of Computer Assisted Tomography
Suez Canal University (EG), Frontier Science Foundation (CH), Gifu University Hospital (JP), Gifu University (JP)
Openalex Percentile: Top 10%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.