Fully 3D-Printed Analytical Device Based on a Novel Floating Electrode Mechanism for Sweat Rate Acquisition

We present herein a unique analytical methodology based on a one-step, three-dimensional printed sensing device for measuring sweat rate noninvasively, which allows for the assessment of hydration levels. It comprises a multilayered construction, and diverse designs are investigated, with the printing components consisting of conductive (CB-PLA) and nonconductive polylactic acid (PLA) printable materials. A prominent aspect is the integration of what is termed here as “floating electrodes”, which advantageously modify the overall impedance of the device during sweat flow measurements. To rationalize the mechanism behind this, multiple sensor configurations, including two-, four-, six-, eight-, or multilayer structures, were thoroughly examined. Although multilayer configurations demonstrated the potential to enhance the overall capacity and sensing area, multiplying the number of electrodes in the device effectively resulted in a higher frequency of impedance variations while providing additional sweat rate data points from a single device. In off-body testing, the optimal system (based on four electrodes) exhibits a calibration range of 1–10 μL min–1 with a total capacity of 37 μL and a correlation coefficient of r = 0.927 between the sensor and timer-based flow validation. With variations oscillating between 0 and 15% as well as a strong correlation (r = 0.79–0.93) with the cotton patch gravimetric method, on-body validation utilizing iontophoresis and cycling to generate sweat in the tested subjects demonstrated strong correlation with standard methods. Our findings illustrate the potential of the “floating electrode” concept for scaling up to larger and higher-capacity sweat sensing devices, as it functions effectively in both single- and double-layer designs. In addition, 3D printing technology will allow for on-demand customization of the corresponding analytical devices.

Authors

Institutions

Publication Details

Journal
ACS Measurement Science Au
Published
2026-05-10
DOI
https://doi.org/10.1021/acsmeasuresciau.6c00056
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fully 3D-Printed Analytical Device Based on a Novel Floating Electrode Mechanism for Sweat Rate Acquisition

Silvia Pérez-Piñero, Vicente Ávila-Gandía, Xing Xuan, Isabel Maria Diaz Lozano et al.
ACS Measurement Science Au
Advanced Sensor and Energy Harvesting Materials
article

Fully 3D-Printed Analytical Device Based on a Novel Floating Electrode Mechanism for Sweat Rate Acquisition

Silvia Pérez-Piñero, Vicente Ávila-Gandía, Xing Xuan, Isabel Maria Diaz Lozano, María Cuartero, Daniel Rojas, Gastón A. Crespo
article en

Abstract

We present herein a unique analytical methodology based on a one-step, three-dimensional printed sensing device for measuring sweat rate noninvasively, which allows for the assessment of hydration levels. It comprises a multilayered construction, and diverse designs are investigated, with the printing components consisting of conductive (CB-PLA) and nonconductive polylactic acid (PLA) printable materials. A prominent aspect is the integration of what is termed here as “floating electrodes”, which advantageously modify the overall impedance of the device during sweat flow measurements. To rationalize the mechanism behind this, multiple sensor configurations, including two-, four-, six-, eight-, or multilayer structures, were thoroughly examined. Although multilayer configurations demonstrated the potential to enhance the overall capacity and sensing area, multiplying the number of electrodes in the device effectively resulted in a higher frequency of impedance variations while providing additional sweat rate data points from a single device. In off-body testing, the optimal system (based on four electrodes) exhibits a calibration range of 1–10 μL min–1 with a total capacity of 37 μL and a correlation coefficient of r = 0.927 between the sensor and timer-based flow validation. With variations oscillating between 0 and 15% as well as a strong correlation (r = 0.79–0.93) with the cotton patch gravimetric method, on-body validation utilizing iontophoresis and cycling to generate sweat in the tested subjects demonstrated strong correlation with standard methods. Our findings illustrate the potential of the “floating electrode” concept for scaling up to larger and higher-capacity sweat sensing devices, as it functions effectively in both single- and double-layer designs. In addition, 3D printing technology will allow for on-demand customization of the corresponding analytical devices.

ACS Measurement Science Au
Universidad Católica de Murcia (ES), KTH Royal Institute of Technology (SE)
Openalex Percentile: Top 8%
Advanced Sensor and Energy Harvesting Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.