Clinical Effective Dynamic Range and the Measurement Floor of SITA-Faster Visual Field Tests

PURPOSE: To describe the effective dynamic range (defined below) using the Swedish Interactive Thresholding Algorithm (SITA)-Faster perimetry and the effects of reliability. METHODS: Nine thousand six hundred and seventy-two test-retest pairs of the 52 test locations of the 24-2 SITA-Faster visual field tests from 1468 eyes of 748 subjects were analysed. The outcome measures were the number of discernible clinical steps, breakpoints and the measurement floor. Effective dynamic range was first assessed using an established approach of smoothed Loess functions (Method 1) applied to test-retest sensitivity pairs. Method 1 was a geometric approach for estimating the outcome measures. Method 2 used the same Loess functions applied to the data in Bland-Altman form. In Method 3, Gaussian smoothing and multisegmental linear regression were applied to the standard deviation of test-retest pairs as a function of mean sensitivity. Analyses were performed on the total cohort, a false positive rate ≤15% subgroup and a false positive rate 0% subgroup. RESULTS: Method 1 returned four clinical steps with intervals bounded by 37 dB, 27 dB and 20-21 dB (the measurement floor), similar across each reliability condition. Method 2 returned the same number of clinical steps, but with a lower measurement floor (18-19 dB). Method 3 led to one to two additional steps and a floor of 18-21 dB. In general, using a false positive rate of 0% resulted in a relatively lower measurement floor value and more breakpoints in comparison with higher false positive rate tolerances. CONCLUSIONS: Using a conventional reliability criterion of false positive rate ≤15%, SITA-Faster has four to five meaningful clinical steps/intervals and a measurement floor of 18-21 dB, which is slightly higher than estimates from SITA-Standard (likely attributable to the higher sensitivity values and greater variability of SITA-Faster). Below this level, clinicians should consider other perimetric approaches, such as 10-2 and/or testing with a Goldmann size V stimulus.

Authors

Institutions

Publication Details

Journal
Ophthalmic and Physiological Optics
Published
2026-05-11
DOI
https://doi.org/10.1007/s44402-026-00103-6
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Effective Dynamic Range and the Measurement Floor of SITA-Faster Visual Field Tests

Michael Kalloniatis, Henrietta Wang, Jack Phu
Ophthalmic and Physiological Optics
Glaucoma and retinal disorders
article

Clinical Effective Dynamic Range and the Measurement Floor of SITA-Faster Visual Field Tests

Michael Kalloniatis, Henrietta Wang, Jack Phu
article en

Abstract

PURPOSE: To describe the effective dynamic range (defined below) using the Swedish Interactive Thresholding Algorithm (SITA)-Faster perimetry and the effects of reliability. METHODS: Nine thousand six hundred and seventy-two test-retest pairs of the 52 test locations of the 24-2 SITA-Faster visual field tests from 1468 eyes of 748 subjects were analysed. The outcome measures were the number of discernible clinical steps, breakpoints and the measurement floor. Effective dynamic range was first assessed using an established approach of smoothed Loess functions (Method 1) applied to test-retest sensitivity pairs. Method 1 was a geometric approach for estimating the outcome measures. Method 2 used the same Loess functions applied to the data in Bland-Altman form. In Method 3, Gaussian smoothing and multisegmental linear regression were applied to the standard deviation of test-retest pairs as a function of mean sensitivity. Analyses were performed on the total cohort, a false positive rate ≤15% subgroup and a false positive rate 0% subgroup. RESULTS: Method 1 returned four clinical steps with intervals bounded by 37 dB, 27 dB and 20-21 dB (the measurement floor), similar across each reliability condition. Method 2 returned the same number of clinical steps, but with a lower measurement floor (18-19 dB). Method 3 led to one to two additional steps and a floor of 18-21 dB. In general, using a false positive rate of 0% resulted in a relatively lower measurement floor value and more breakpoints in comparison with higher false positive rate tolerances. CONCLUSIONS: Using a conventional reliability criterion of false positive rate ≤15%, SITA-Faster has four to five meaningful clinical steps/intervals and a measurement floor of 18-21 dB, which is slightly higher than estimates from SITA-Standard (likely attributable to the higher sensitivity values and greater variability of SITA-Faster). Below this level, clinicians should consider other perimetric approaches, such as 10-2 and/or testing with a Goldmann size V stimulus.

Ophthalmic and Physiological Optics
Deakin University (AU), UNSW Sydney (AU), University of Houston (US)
National Health and Medical Research Council
Openalex Percentile: Top 6%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.