Integrating measurement and simulation for evaluation of patient‐specific whole‐body dosimetry in clinical BNCT

BACKGROUND: Boron neutron capture therapy (BNCT) is a treatment modality that utilises high intensity neutron beam in combination with a boron drug to target cancer cells at the cellular level. Despite decades of research in this field, there are very few reports on out-of-field patient dose of BNCT. PURPOSE: To quantify patient whole-body out-of-field dose during clinical BNCT for recurrent head-and-neck cancer using an integrated measurement-simulation workflow. METHODS: The data of 271 patients that received BNCT for recurrent head and neck were analyzed for this study.First, the neutron and gamma ray doses were measured using activation foils and thermoluminescent dosimeters, respectively. The detectors were placed onto the surface of the patient at various locations and the reaction rate of the metal foils were measured. Second, PHITS Monte Carlo simulation was performed to evaluate the neutron energy spectrum at each location and the corresponding neutron equivalent dose was determined. The total dose was calculated by summing the neutron equivalent dose and the gamma ray dose. The results were compared with other radiotherapy modalities. RESULTS: The mean ± standard deviation of equivalent dose at neck, chest, abdomen, waist, knee, and ankle were calculated to be 1.93 ± 1.12, 0.71 ± 0.42, 0.23 ± 0.12, 0.11 ± 0.06, 0.05 ± 0.03, 0.02 ± 0.01 Gy-eq, respectively. Excluding the neck region, the gamma ray dose was dominant at all measurement points. From the abdomen below, dose from the primary gamma ray were dominant, indicating a potential for reducing the dose to these regions by installing additional lead shielding in the treatment room walls. In comparison to other radiotherapy modalities, out-of-field dose of BNCT was found to be similar. CONCLUSION: The whole-body dose for patients that received BNCT for recurrent head and neck cancer were estimated by scaling PHITS derived neutron dose using multifoil reaction rates and adding measured gamma dose. The results showed the out-of-field dose was comparable to other radiotherapy modalities and further added confidence that BNCT is a safe treatment modality.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-05-01
DOI
https://doi.org/10.1002/mp.70474
Primary Topic
Boron Compounds in Chemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integrating measurement and simulation for evaluation of patient‐specific whole‐body dosimetry in clinical BNCT

Naonori Hu, Ryo Kakino, Nishiki Matsubayashi, Keiji Nihei et al.
Medical Physics
Boron Compounds in Chemistry
article

Integrating measurement and simulation for evaluation of patient‐specific whole‐body dosimetry in clinical BNCT

Naonori Hu, Ryo Kakino, Nishiki Matsubayashi, Keiji Nihei, Mai Nojiri, Teruhito Aihara, Takushi Takata, Yuki Yoshino, Satoshi Takeno, Hiroki Tanaka, Akinori Sasaki, Koji Ono
article en

Abstract

BACKGROUND: Boron neutron capture therapy (BNCT) is a treatment modality that utilises high intensity neutron beam in combination with a boron drug to target cancer cells at the cellular level. Despite decades of research in this field, there are very few reports on out-of-field patient dose of BNCT. PURPOSE: To quantify patient whole-body out-of-field dose during clinical BNCT for recurrent head-and-neck cancer using an integrated measurement-simulation workflow. METHODS: The data of 271 patients that received BNCT for recurrent head and neck were analyzed for this study.First, the neutron and gamma ray doses were measured using activation foils and thermoluminescent dosimeters, respectively. The detectors were placed onto the surface of the patient at various locations and the reaction rate of the metal foils were measured. Second, PHITS Monte Carlo simulation was performed to evaluate the neutron energy spectrum at each location and the corresponding neutron equivalent dose was determined. The total dose was calculated by summing the neutron equivalent dose and the gamma ray dose. The results were compared with other radiotherapy modalities. RESULTS: The mean ± standard deviation of equivalent dose at neck, chest, abdomen, waist, knee, and ankle were calculated to be 1.93 ± 1.12, 0.71 ± 0.42, 0.23 ± 0.12, 0.11 ± 0.06, 0.05 ± 0.03, 0.02 ± 0.01 Gy-eq, respectively. Excluding the neck region, the gamma ray dose was dominant at all measurement points. From the abdomen below, dose from the primary gamma ray were dominant, indicating a potential for reducing the dose to these regions by installing additional lead shielding in the treatment room walls. In comparison to other radiotherapy modalities, out-of-field dose of BNCT was found to be similar. CONCLUSION: The whole-body dose for patients that received BNCT for recurrent head and neck cancer were estimated by scaling PHITS derived neutron dose using multifoil reaction rates and adding measured gamma dose. The results showed the out-of-field dose was comparable to other radiotherapy modalities and further added confidence that BNCT is a safe treatment modality.

Medical PhysicsVol. 53(5)
Kyoto Prefectural University of Medicine (JP), Kyocera (Japan) (JP), Osaka University of Pharmaceutical Sciences (JP)
Good health and well-being
Openalex Percentile: Top 6%
Boron Compounds in Chemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.