Mitigation of ionization‐induced quantitative bias in <scp>LC</scp> – <scp>MS</scp> / <scp>MS</scp> analysis of <i>N</i> ‐nitrosamine drug‐substance‐related impurities in <scp>Rasagiline</scp> pharmaceutical products

Abstract Accurate quantification of N ‐nitrosamine drug‐substance–related impurities (NDSRIs) in pharmaceutical products requires careful control of ionization‐induced artifacts during LC–MS/MS analysis. In pharmaceutical matrices where the active pharmaceutical ingredient (API) is present at concentrations several orders of magnitude higher than the target impurity, in‐source fragmentation can generate overlapping product ions, leading to significant quantitative bias. In this study, we demonstrate that certain atmospheric‐pressure ionization conditions induce in‐source decomposition of both rasagiline and its corresponding nitroso impurity, producing a common fragment ion at m / z 117. This overlap can compromise analytical specificity and potentially result in overestimation of impurity levels. To mitigate this ionization‐induced quantitative bias, Heated Electrospray Ionization (H‐ESI) parameters were systematically optimized to preserve the intact protonated molecular ion ( m / z 201) while minimizing thermally induced fragmentation. In addition, a gradient LC method incorporating a high‐organic wash step and flow diversion strategy was established to enhance analytical robustness under routine quality control conditions. The optimized LC–MS/MS method demonstrated excellent linearity ( R 2 &gt; 0.998), acceptable precision (RSD ≤ 11.5%), and sufficient sensitivity (LOQ 30 ng/mL) to meet current regulatory requirements. These findings highlight the critical importance of ion‐source selection in nitrosamine impurity analysis and provide a practical strategy to reduce the risk of ionization‐induced quantitative bias in pharmaceutical LC–MS/MS applications.

Authors

Institutions

Publication Details

Journal
Bulletin of the Korean Chemical Society
Published
2026-05-09
DOI
https://doi.org/10.1002/bkcs.70171
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mitigation of ionization‐induced quantitative bias in LC – MS / MS analysis of N ‐nitrosamine drug‐substance‐related impurities in Rasagiline pharmaceutical products

Han Bin Oh, Chan Hong Min
Bulletin of the Korean Chemical Society
Mass Spectrometry Techniques and Applications
article

Mitigation of ionization‐induced quantitative bias in LC – MS / MS analysis of N ‐nitrosamine drug‐substance‐related impurities in Rasagiline pharmaceutical products

Han Bin Oh, Chan Hong Min
article en

Abstract

Abstract Accurate quantification of N ‐nitrosamine drug‐substance–related impurities (NDSRIs) in pharmaceutical products requires careful control of ionization‐induced artifacts during LC–MS/MS analysis. In pharmaceutical matrices where the active pharmaceutical ingredient (API) is present at concentrations several orders of magnitude higher than the target impurity, in‐source fragmentation can generate overlapping product ions, leading to significant quantitative bias. In this study, we demonstrate that certain atmospheric‐pressure ionization conditions induce in‐source decomposition of both rasagiline and its corresponding nitroso impurity, producing a common fragment ion at m / z 117. This overlap can compromise analytical specificity and potentially result in overestimation of impurity levels. To mitigate this ionization‐induced quantitative bias, Heated Electrospray Ionization (H‐ESI) parameters were systematically optimized to preserve the intact protonated molecular ion ( m / z 201) while minimizing thermally induced fragmentation. In addition, a gradient LC method incorporating a high‐organic wash step and flow diversion strategy was established to enhance analytical robustness under routine quality control conditions. The optimized LC–MS/MS method demonstrated excellent linearity ( R 2 &gt; 0.998), acceptable precision (RSD ≤ 11.5%), and sufficient sensitivity (LOQ 30 ng/mL) to meet current regulatory requirements. These findings highlight the critical importance of ion‐source selection in nitrosamine impurity analysis and provide a practical strategy to reduce the risk of ionization‐induced quantitative bias in pharmaceutical LC–MS/MS applications.

Bulletin of the Korean Chemical Society
Sogang University (KR)
Ministry of Education, Ministry of Food and Drug Safety, Korea Health Industry Development Institute
Openalex Percentile: Top 10%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.