Commissioning of a 10 MV photon beam from an elekta synergy agility linear accelerator in the eclipse treatment planning system: A dosimetric comparison of AAA and Acuros XB

Abstract Purpose Published end-to-end validation data for 10 MV photon beams in mixed-vendor radiotherapy environments remain limited. A 10 MV photon beam from an Elekta Synergy Agility linear accelerator was commissioned in Eclipse (version 17.1), and the dosimetric performance of AAA and Acuros XB was evaluated using end-to-end testing. Methods Beam data were measured in a PTW Beamscan water phantom and then imported into Eclipse to generate beam models for AAA and AXB. Eclipse was integrated with the Mosaiq oncology information system to enable better workflow. Following IAEA TECDOC-1583 guidelines, end-to-end test plans were created in Eclipse and delivered to a CIRS thorax phantom (model 002LFC) containing soft-tissue, lung, and bone-equivalent inserts. Absolute dose measurements were compared for AAA, AXB dose-to-water (Dw), and AXB dose-to-medium (Dm), with stopping-power corrections applied for lung and bone regions. Results End-to-end testing demonstrated good overall agreement for both algorithms. AAA showed consistent performance in soft tissue and closer agreement in bone-equivalent regions, while AXB performed better in lung-equivalent regions. Larger deviations were observed for AXB in posterior beam arrangements where the beam passed through bone before reaching the target. Conclusion The 10 MV photon beam was successfully commissioned and verified for clinical use in the Eclipse system. Both AAA and AXB demonstrated clinically acceptable accuracy for 3DCRT. Although algorithm-dependent differences were observed in heterogeneous conditions, AAA remains suitable for 3DCRT planning when AXB is unavailable. This work contributes new end-to-end validation data for a 10 MV Elekta-Eclipse configuration, addressing a current gap in the literature for mixed-vendor radiotherapy systems.

Authors

Institutions

Publication Details

Journal
Health and Technology
Published
2026-05-09
DOI
https://doi.org/10.1007/s12553-026-01079-6
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Commissioning of a 10 MV photon beam from an elekta synergy agility linear accelerator in the eclipse treatment planning system: A dosimetric comparison of AAA and Acuros XB

Chanté Jooste, Christoph Trauernicht, Christoffel Jacobus van Reenen, Andrea Marais et al.
Health and Technology
Advanced Radiotherapy Techniques
article

Commissioning of a 10 MV photon beam from an elekta synergy agility linear accelerator in the eclipse treatment planning system: A dosimetric comparison of AAA and Acuros XB

Chanté Jooste, Christoph Trauernicht, Christoffel Jacobus van Reenen, Andrea Marais, Tamzyn Payne, Jessica Ester Olivier, Mohlapoli Mohlapholi, Jake Henry Blank
article en

Abstract

Abstract Purpose Published end-to-end validation data for 10 MV photon beams in mixed-vendor radiotherapy environments remain limited. A 10 MV photon beam from an Elekta Synergy Agility linear accelerator was commissioned in Eclipse (version 17.1), and the dosimetric performance of AAA and Acuros XB was evaluated using end-to-end testing. Methods Beam data were measured in a PTW Beamscan water phantom and then imported into Eclipse to generate beam models for AAA and AXB. Eclipse was integrated with the Mosaiq oncology information system to enable better workflow. Following IAEA TECDOC-1583 guidelines, end-to-end test plans were created in Eclipse and delivered to a CIRS thorax phantom (model 002LFC) containing soft-tissue, lung, and bone-equivalent inserts. Absolute dose measurements were compared for AAA, AXB dose-to-water (Dw), and AXB dose-to-medium (Dm), with stopping-power corrections applied for lung and bone regions. Results End-to-end testing demonstrated good overall agreement for both algorithms. AAA showed consistent performance in soft tissue and closer agreement in bone-equivalent regions, while AXB performed better in lung-equivalent regions. Larger deviations were observed for AXB in posterior beam arrangements where the beam passed through bone before reaching the target. Conclusion The 10 MV photon beam was successfully commissioned and verified for clinical use in the Eclipse system. Both AAA and AXB demonstrated clinically acceptable accuracy for 3DCRT. Although algorithm-dependent differences were observed in heterogeneous conditions, AAA remains suitable for 3DCRT planning when AXB is unavailable. This work contributes new end-to-end validation data for a 10 MV Elekta-Eclipse configuration, addressing a current gap in the literature for mixed-vendor radiotherapy systems.

Health and Technology
Stellenbosch University (ZA), Tygerberg Hospital (ZA)
Universiteit Stellenbosch
Openalex Percentile: Top 6%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.