Reliability and validity of a smartwatch with a photoplethysmograph: Comparison with a Holter electrocardiograph

Although consumer-grade smartwatches with a photoplethysmograph (PPG) can measure heart rate, their reliability and validity in healthy participants under free-living conditions remain unclear. Therefore, this study compared the heart rates measured by a smartwatch with PPG and a clinically accepted Holter electrocardiograph in healthy adult participants under completely free-living conditions. Ten participants wore a Holter recorder on their left chest and a Garmin vivosmart 5 on their non-dominant wrist simultaneously for 72-96 h. Averages were calculated every 2 min from the data obtained, and timestamps were used for alignment. Agreement between the measurements taken by the two devices was verified using intraclass correlation and Bland-Altman analyses. Error trends for Garmin vivosmart 5 were examined using linear regression analysis. The overall intraclass correlation coefficient range was 0.819-0.937 (mean: 0.902), indicating strong agreement. The mean bias was 1.16 bpm, and the mean limit of agreement was ±12.4 (7.87-20.6) bpm. Furthermore, the linear regression line intercept was negative for all participants, indicating that Garmin vivosmart 5 tended to underestimate the heart rate; however, the slope was positive and close to 0. Overall, the smartwatch maintained a certain accuracy level under fluctuating heart rates and demonstrated reasonable reliability during sleep or daily activities.

Authors

Institutions

Publication Details

Journal
Chronobiology International
Published
2026-05-08
DOI
https://doi.org/10.1080/07420528.2026.2669593
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reliability and validity of a smartwatch with a photoplethysmograph: Comparison with a Holter electrocardiograph

Nobue Arimoto, Tomoya Tamei, Hideyuki Shiotani, Chieko Tan et al.
Chronobiology International
Non-Invasive Vital Sign Monitoring
article

Reliability and validity of a smartwatch with a photoplethysmograph: Comparison with a Holter electrocardiograph

Nobue Arimoto, Tomoya Tamei, Hideyuki Shiotani, Chieko Tan, Atsue Ishii
article en

Abstract

Although consumer-grade smartwatches with a photoplethysmograph (PPG) can measure heart rate, their reliability and validity in healthy participants under free-living conditions remain unclear. Therefore, this study compared the heart rates measured by a smartwatch with PPG and a clinically accepted Holter electrocardiograph in healthy adult participants under completely free-living conditions. Ten participants wore a Holter recorder on their left chest and a Garmin vivosmart 5 on their non-dominant wrist simultaneously for 72-96 h. Averages were calculated every 2 min from the data obtained, and timestamps were used for alignment. Agreement between the measurements taken by the two devices was verified using intraclass correlation and Bland-Altman analyses. Error trends for Garmin vivosmart 5 were examined using linear regression analysis. The overall intraclass correlation coefficient range was 0.819-0.937 (mean: 0.902), indicating strong agreement. The mean bias was 1.16 bpm, and the mean limit of agreement was ±12.4 (7.87-20.6) bpm. Furthermore, the linear regression line intercept was negative for all participants, indicating that Garmin vivosmart 5 tended to underestimate the heart rate; however, the slope was positive and close to 0. Overall, the smartwatch maintained a certain accuracy level under fluctuating heart rates and demonstrated reasonable reliability during sleep or daily activities.

Chronobiology International
Kobe Tokiwa University (JP), Kobe University (JP), Nara Institute of Science and Technology (JP)
Openalex Percentile: Top 8%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.