Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: A Pilot Study

Abstract: Background and aims: Accurate measurement of colorectal polyp size is important for planning resection and post-polypectomy surveillance. This study evaluated a through-the-scope laser-based probe for polyp sizing during colonoscopy. Methods: A prospective pilot study was conducted at CHUM, where polyps were measured using a laser-based probe. The primary outcome was the accuracy of the laser-based probe, defined as mean bias with microscopic sizing as the reference standard. Secondary outcomes included agreement with the reference, implementation rate, technical success rate, reasons for failure, and per-polyp measurement time. Results: Among 323 patients, 402 colorectal polyps were detected. Laser-based measurement was attempted in 237 polyps (59.0% 237/402; 95% CI: 54.0–63.8), of which 149 had microscopic reference measurements. The technical success rate was 40.9% (97/237; 95% CI, 34.6–47.5), with a mean measurement time of 1:35. Accounting for non-en-bloc resections, 41 of 149 polyps (27.5%; 95% CI, 20.7–35.5) had paired laser–microscopy measurements for the primary analysis. Laser-based measurement showed minimal bias (−0.10 mm; 95% CI, −0.30 to 0.10; P = 0.328), meeting the prespecified noninferiority margin of 0.4 mm. Agreement with the reference standard was high (ICC = 0.91; 95% CI, 0.83–0.95), with near-perfect categorical concordance (κ = 0.85) and 95.1% classification accuracy (39/41; 95% CI, 82.2–99.2). Bland–Altman limits of agreement ranged from −1.34 mm to 1.14 mm. Conclusions: Laser-based measurement provided accurate polyp sizing when obtained. However, technical success was limited by longer measurement times and device issues.

Authors

Institutions

Publication Details

Journal
Endoscopy International Open
Published
2026-05-08
DOI
https://doi.org/10.1055/a-2872-7167
Primary Topic
Colorectal Cancer Screening and Detection
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: A Pilot Study

Roupen Djinbachian, Dane Christina Daoud, Simon Bouchard, Victoire Michal et al.
Endoscopy International Open
Colorectal Cancer Screening and Detection
article

Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: A Pilot Study

Roupen Djinbachian, Dane Christina Daoud, Simon Bouchard, Victoire Michal, Jeremy Liu Chen Kiow, Preslava Aleksieva, Megan Oleksiw, Eric Cristea, Mohamed Zineddine Mahdadi, M Bouin, R Battat, B Panzini, Daniel von Renteln, Claire Gefflot
article en

Abstract

Abstract: Background and aims: Accurate measurement of colorectal polyp size is important for planning resection and post-polypectomy surveillance. This study evaluated a through-the-scope laser-based probe for polyp sizing during colonoscopy. Methods: A prospective pilot study was conducted at CHUM, where polyps were measured using a laser-based probe. The primary outcome was the accuracy of the laser-based probe, defined as mean bias with microscopic sizing as the reference standard. Secondary outcomes included agreement with the reference, implementation rate, technical success rate, reasons for failure, and per-polyp measurement time. Results: Among 323 patients, 402 colorectal polyps were detected. Laser-based measurement was attempted in 237 polyps (59.0% 237/402; 95% CI: 54.0–63.8), of which 149 had microscopic reference measurements. The technical success rate was 40.9% (97/237; 95% CI, 34.6–47.5), with a mean measurement time of 1:35. Accounting for non-en-bloc resections, 41 of 149 polyps (27.5%; 95% CI, 20.7–35.5) had paired laser–microscopy measurements for the primary analysis. Laser-based measurement showed minimal bias (−0.10 mm; 95% CI, −0.30 to 0.10; P = 0.328), meeting the prespecified noninferiority margin of 0.4 mm. Agreement with the reference standard was high (ICC = 0.91; 95% CI, 0.83–0.95), with near-perfect categorical concordance (κ = 0.85) and 95.1% classification accuracy (39/41; 95% CI, 82.2–99.2). Bland–Altman limits of agreement ranged from −1.34 mm to 1.14 mm. Conclusions: Laser-based measurement provided accurate polyp sizing when obtained. However, technical success was limited by longer measurement times and device issues.

Endoscopy International Open
McGill University Health Centre (CA), Centre Hospitalier de l’Université de Montréal (CA), Université de Montréal (CA)
Boston Scientific Corporation
Openalex Percentile: Top 7%
Colorectal Cancer Screening and Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.