Effect of output intensity of light curing unit on coronal tooth discoloration induced by silicate calcium-based materials: an ex vivo study

Objective: Clinicians have various curing options due to new light-cure devices with different intensities. The present study aimed to evaluate the effect of two light-curing intensities (1,200 mW/cm² vs. 3,000 mW/cm²) on coronal discoloration induced by two calcium silicate-based cements (OrthoMTA and RetroMTA) in an endodontically treated bovine tooth model. Materials and Methods: Forty single-rooted bovine incisors were prepared, and their canals were filled with blood-soaked foam to the cemento-enamel junction. The samples were randomly divided into four groups (n = 10 per group) based on the cement used (OrthoMTA or RetroMTA) and the subsequent light-curing intensity. For all groups, a 2-mm layer of the assigned cement was placed. This was covered by 1 mm of resin-modified glass ionomer (RMGI), a bonding agent, and 2 mm of composite. In two groups, all light-curing steps (RMGI, bonding agent, composite) were performed at 1,200 mW/cm². In the other two groups, curing was performed at 3,000 mW/cm². All samples underwent 5,000 cycles of thermocycling. Color was assessed with a spectrophotometer immediately after composite placement and after aging. Color change (ΔE) and lightness change (ΔL) were calculated. Results: All groups exhibited ΔE values exceeding the clinically acceptable threshold (ΔE > 3.3) and demonstrated a decrease in ΔL after aging. Statistical analysis (two-way ANOVA, α = 0.05) revealed no significant difference in ΔE or ΔL between samples cured with high-intensity (3,000 mW/cm²) or low-intensity (1,200 mW/cm²) light, regardless of the cement type used (p > 0.05). Conclusions: Within the limitations of this study, the same amount of energy produced by high- or low-intensity light-cures (1,200 or 3,000 mW/cm²), when delivering the same total energy, did not result in a statistically significant difference in coronal discoloration caused by OrthoMTA or RetroMTA.

Authors

Institutions

Publication Details

Journal
Biomaterial Investigations in Dentistry
Published
2026-05-08
DOI
https://doi.org/10.2340/biid.v13.45896
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of output intensity of light curing unit on coronal tooth discoloration induced by silicate calcium-based materials: an ex vivo study

Noushin Shokouhinejad, Fatemeh Farshad, Parham Pedram, Mohammad Javad KharaziFard et al.
Biomaterial Investigations in Dentistry
Dental materials and restorations
article

Effect of output intensity of light curing unit on coronal tooth discoloration induced by silicate calcium-based materials: an ex vivo study

Noushin Shokouhinejad, Fatemeh Farshad, Parham Pedram, Mohammad Javad KharaziFard, Elham Ahmadi
article en

Abstract

Objective: Clinicians have various curing options due to new light-cure devices with different intensities. The present study aimed to evaluate the effect of two light-curing intensities (1,200 mW/cm² vs. 3,000 mW/cm²) on coronal discoloration induced by two calcium silicate-based cements (OrthoMTA and RetroMTA) in an endodontically treated bovine tooth model. Materials and Methods: Forty single-rooted bovine incisors were prepared, and their canals were filled with blood-soaked foam to the cemento-enamel junction. The samples were randomly divided into four groups (n = 10 per group) based on the cement used (OrthoMTA or RetroMTA) and the subsequent light-curing intensity. For all groups, a 2-mm layer of the assigned cement was placed. This was covered by 1 mm of resin-modified glass ionomer (RMGI), a bonding agent, and 2 mm of composite. In two groups, all light-curing steps (RMGI, bonding agent, composite) were performed at 1,200 mW/cm². In the other two groups, curing was performed at 3,000 mW/cm². All samples underwent 5,000 cycles of thermocycling. Color was assessed with a spectrophotometer immediately after composite placement and after aging. Color change (ΔE) and lightness change (ΔL) were calculated. Results: All groups exhibited ΔE values exceeding the clinically acceptable threshold (ΔE > 3.3) and demonstrated a decrease in ΔL after aging. Statistical analysis (two-way ANOVA, α = 0.05) revealed no significant difference in ΔE or ΔL between samples cured with high-intensity (3,000 mW/cm²) or low-intensity (1,200 mW/cm²) light, regardless of the cement type used (p > 0.05). Conclusions: Within the limitations of this study, the same amount of energy produced by high- or low-intensity light-cures (1,200 or 3,000 mW/cm²), when delivering the same total energy, did not result in a statistically significant difference in coronal discoloration caused by OrthoMTA or RetroMTA.

Biomaterial Investigations in DentistryVol. 13
Shahid Beheshti University of Medical Sciences (IR), Tehran University of Medical Sciences (IR)
Tehran University of Medical Sciences and Health Services
Affordable and clean energy
Openalex Percentile: Top 6%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.