High-Frequency Ultrasonography for Ex Vivo Periodontal Tissue Measurement: Accuracy and Reproducibility Compared to Cone Beam Computed Tomography

OBJECTIVE: The aim of this study was to assess the accuracy and reproducibility of periodontal tissue measurements obtained through ultrasonographic imaging, by comparing them ex vivo to reference measurements acquired via cone beam computed tomography (CBCT). METHODS: A total of 12 hemi-jaws were included in the study. CBCT acquisition was realised, and then ultrasound measurements of periodontal structures were performed. Ultrasound imaging was performed with a linear array 20 MHz probe (field of view 14.1 × 6.4 mm; penetration depth ≈ 4.9 mm; axial/lateral resolution 80/120 µm). The ultrasound images were analysed using ImageJ software. Reference points were made allowing the measurement of the different tissues. A Kolmogorov-Smirnov test and Lin's concordance test were performed for each of the samples due to the size of their population (>50). RESULTS: A total of 348 manual probing measurements were performed. Lin's concordance correlation coefficients were: gingival thickness: 0.479 (95% CI [0.393-0.558]); distance from the alveolar crest to the CEJ: 0.872 (95% CI [0.842-0.897]); distance from the alveolar crest to the gingival margin: 0.794 (95% CI [0.749-0.832]); alveolar bone thickness: -0.0165 (95% CI [-0.0512 to 0.0183]). CONCLUSION: High-frequency ultrasonography demonstrates high accuracy and reproducibility for measuring periodontal structures ex vivo, particularly for the distance from the alveolar crest to the CEJ, but further clinical studies are needed to validate its use in vivo. CLINICAL RELEVANCE: This non-ionising technique offers a safe alternative to CBCT for periodontal assessment, with potential to enhance early diagnosis and monitoring in clinical practice while minimising radiation exposure.

Authors

Institutions

Publication Details

Journal
Ultrasound in Medicine & Biology
Published
2026-05-01
DOI
https://doi.org/10.1016/j.ultrasmedbio.2026.02.013
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-Frequency Ultrasonography for Ex Vivo Periodontal Tissue Measurement: Accuracy and Reproducibility Compared to Cone Beam Computed Tomography

Gaël Y. Rochefort, M. Renaud, Laurent Estrade, Alexis Delpierre et al.
Ultrasound in Medicine & Biology
Dental Radiography and Imaging
article

High-Frequency Ultrasonography for Ex Vivo Periodontal Tissue Measurement: Accuracy and Reproducibility Compared to Cone Beam Computed Tomography

Gaël Y. Rochefort, M. Renaud, Laurent Estrade, Alexis Delpierre, Baptiste Ait Slimane, Victor Rimbaud, Frédéric Denis
article en

Abstract

OBJECTIVE: The aim of this study was to assess the accuracy and reproducibility of periodontal tissue measurements obtained through ultrasonographic imaging, by comparing them ex vivo to reference measurements acquired via cone beam computed tomography (CBCT). METHODS: A total of 12 hemi-jaws were included in the study. CBCT acquisition was realised, and then ultrasound measurements of periodontal structures were performed. Ultrasound imaging was performed with a linear array 20 MHz probe (field of view 14.1 × 6.4 mm; penetration depth ≈ 4.9 mm; axial/lateral resolution 80/120 µm). The ultrasound images were analysed using ImageJ software. Reference points were made allowing the measurement of the different tissues. A Kolmogorov-Smirnov test and Lin's concordance test were performed for each of the samples due to the size of their population (>50). RESULTS: A total of 348 manual probing measurements were performed. Lin's concordance correlation coefficients were: gingival thickness: 0.479 (95% CI [0.393-0.558]); distance from the alveolar crest to the CEJ: 0.872 (95% CI [0.842-0.897]); distance from the alveolar crest to the gingival margin: 0.794 (95% CI [0.749-0.832]); alveolar bone thickness: -0.0165 (95% CI [-0.0512 to 0.0183]). CONCLUSION: High-frequency ultrasonography demonstrates high accuracy and reproducibility for measuring periodontal structures ex vivo, particularly for the distance from the alveolar crest to the CEJ, but further clinical studies are needed to validate its use in vivo. CLINICAL RELEVANCE: This non-ionising technique offers a safe alternative to CBCT for periodontal assessment, with potential to enhance early diagnosis and monitoring in clinical practice while minimising radiation exposure.

Ultrasound in Medicine & Biology
Université de Tours (FR), Inserm (FR), Centre Hospitalier Universitaire de Tours (FR)
Openalex Percentile: Top 5%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.