Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging

OBJECTIVE: To accurately monitor temperature during cryoablation, a minimally invasive technique that destroys tissue locally by forming an ice ball around an inserted cryoprobe.
Approach: We present a dense slowness-shift imaging method that estimates local speed-of-sound changes from ultrasound B-mode images using optical flow. This single-transducer, image-based approach enables mapping of spatial temperature change without requiring additional hardware. Cryoablation experiments were conducted in a tissue-mimicking phantom and ex vivo turkey breast.
Main results: Slowness deviation increased with decreasing temperature. In the phantom, the dependence was linear (αₐ = -20.70 ηs·m⁻¹C°⁻¹), while in turkey breast it followed an exponential relationship (αₜ = 34.04×exp(0.075(-ΔT)-1) ηs·m⁻¹C°⁻¹). The algorithm detected sub-degree temperature variations and accurately tracked cooling down to -39.4 ± 5.6 °C.
Significance: Accurate temperature monitoring is essential for effective and safe cryoablation. This work demonstrates the feasibility of ultrasound-based, noninvasive temperature monitoring during cryoablation, providing a scalable, real-time alternative to existing invasive or high-cost thermal assessment techniques.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-05-08
DOI
https://doi.org/10.1088/1361-6560/ae6af9
Primary Topic
Photoacoustic and Ultrasonic Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging

Tali Ilovitsh, Tal Grutman, Mike Bismuth, Gaya Lamm
Physics in Medicine and Biology
Photoacoustic and Ultrasonic Imaging
article

Cryoablation temperature monitoring with dense ultrasonic speed-of-sound shift imaging

Tali Ilovitsh, Tal Grutman, Mike Bismuth, Gaya Lamm
article en

Abstract

OBJECTIVE: To accurately monitor temperature during cryoablation, a minimally invasive technique that destroys tissue locally by forming an ice ball around an inserted cryoprobe.
Approach: We present a dense slowness-shift imaging method that estimates local speed-of-sound changes from ultrasound B-mode images using optical flow. This single-transducer, image-based approach enables mapping of spatial temperature change without requiring additional hardware. Cryoablation experiments were conducted in a tissue-mimicking phantom and ex vivo turkey breast.
Main results: Slowness deviation increased with decreasing temperature. In the phantom, the dependence was linear (αₐ = -20.70 ηs·m⁻¹C°⁻¹), while in turkey breast it followed an exponential relationship (αₜ = 34.04×exp(0.075(-ΔT)-1) ηs·m⁻¹C°⁻¹). The algorithm detected sub-degree temperature variations and accurately tracked cooling down to -39.4 ± 5.6 °C.
Significance: Accurate temperature monitoring is essential for effective and safe cryoablation. This work demonstrates the feasibility of ultrasound-based, noninvasive temperature monitoring during cryoablation, providing a scalable, real-time alternative to existing invasive or high-cost thermal assessment techniques.

Physics in Medicine and Biology
Tel Aviv University (IL), Academic College of Tel Aviv-Yafo (IL)
Israel Cancer Research Fund, Israel Science Foundation
Openalex Percentile: Top 10%
Photoacoustic and Ultrasonic Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.