1315 Bent but Not Broken: Hypoglossal Nerve Stimulation Malfunction Caused by Severe Bending of an Intact Sensing Lead

Abstract Introduction Hypoglossal nerve stimulation (HNS) provides an alternate treatment option for CPAP intolerant or non-adherent patients with moderate to severe obstructive sleep apnea. HNS is delivered via a stimulation lead placed around the hypoglossal nerve, an implanted pulse generator, and a respiratory sensing lead placed in the chest. While there have been reports of sensing lead malfunction due to migration of lead or tip fracture, we present a unique case of sensing lead malfunction caused by severe bending of the lead with an intact tip. Report of case(s) A 64-year-old male with severe obstructive sleep apnea underwent HNS implantation due to dissatisfaction with CPAP therapy. Pre-procedure home sleep apnea test showed an AHI-4% of 37.7/hour. Post-operative course was uncomplicated. Initial waveform interrogation showed appropriate respiratory sensing. An HNS fine tune study showed only partial improvement of sleep apnea with stimulation-respiration dyssynchrony. Subsequent wave form analysis demonstrated a shark fin morphology. Waveform inversion was attempted, but an advanced fine tune study showed persistent respiratory events at different amplitude, electrode and timing settings. An awake endoscopy noted that mouth opening and neck flexion were associated with poor airway opening and use of a foam cervical collar was recommended. A home sleep apnea test at the recommended electrode settings with use of a foam cervical collar showed a treatment pAHI-4% of 40.5/hour. Due to persistently elevated AHI with aberrant respiratory sensing despite several setting adjustments, a revision of the sensing lead was pursued. Intraoperatively, the internal component of the sensing lead was noted to be severely bent with an intact lead tip. Following the sensing lead exchange, the shark fin waveform morphology resolved. A repeat home sleep apnea test showed improved treatment pAHI-4% of 13.1/hour. Conclusion A retrospective review of HNS related adverse events published in June 2024 noted that sensing lead malfunction included either lead migration or fracture. Our case highlights yet another cause of malfunction due to kinking of an intact sensing lead. It also demonstrates that HNS with non-physiologic shark-fin waveform may not be effective in improving sleep apnea despite recommended programming changes even with foam cervical collar. Support (if any)

Authors

Institutions

Publication Details

Journal
SLEEP
Published
2026-05-01
DOI
https://doi.org/10.1093/sleep/zsag091.1314
Primary Topic
Obstructive Sleep Apnea Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

1315 Bent but Not Broken: Hypoglossal Nerve Stimulation Malfunction Caused by Severe Bending of an Intact Sensing Lead

Paul Covello, Karthy Thiagarajan, Richard Byrne
SLEEP
Obstructive Sleep Apnea Research
article

1315 Bent but Not Broken: Hypoglossal Nerve Stimulation Malfunction Caused by Severe Bending of an Intact Sensing Lead

Paul Covello, Karthy Thiagarajan, Richard Byrne
article en

Abstract

Abstract Introduction Hypoglossal nerve stimulation (HNS) provides an alternate treatment option for CPAP intolerant or non-adherent patients with moderate to severe obstructive sleep apnea. HNS is delivered via a stimulation lead placed around the hypoglossal nerve, an implanted pulse generator, and a respiratory sensing lead placed in the chest. While there have been reports of sensing lead malfunction due to migration of lead or tip fracture, we present a unique case of sensing lead malfunction caused by severe bending of the lead with an intact tip. Report of case(s) A 64-year-old male with severe obstructive sleep apnea underwent HNS implantation due to dissatisfaction with CPAP therapy. Pre-procedure home sleep apnea test showed an AHI-4% of 37.7/hour. Post-operative course was uncomplicated. Initial waveform interrogation showed appropriate respiratory sensing. An HNS fine tune study showed only partial improvement of sleep apnea with stimulation-respiration dyssynchrony. Subsequent wave form analysis demonstrated a shark fin morphology. Waveform inversion was attempted, but an advanced fine tune study showed persistent respiratory events at different amplitude, electrode and timing settings. An awake endoscopy noted that mouth opening and neck flexion were associated with poor airway opening and use of a foam cervical collar was recommended. A home sleep apnea test at the recommended electrode settings with use of a foam cervical collar showed a treatment pAHI-4% of 40.5/hour. Due to persistently elevated AHI with aberrant respiratory sensing despite several setting adjustments, a revision of the sensing lead was pursued. Intraoperatively, the internal component of the sensing lead was noted to be severely bent with an intact lead tip. Following the sensing lead exchange, the shark fin waveform morphology resolved. A repeat home sleep apnea test showed improved treatment pAHI-4% of 13.1/hour. Conclusion A retrospective review of HNS related adverse events published in June 2024 noted that sensing lead malfunction included either lead migration or fracture. Our case highlights yet another cause of malfunction due to kinking of an intact sensing lead. It also demonstrates that HNS with non-physiologic shark-fin waveform may not be effective in improving sleep apnea despite recommended programming changes even with foam cervical collar. Support (if any)

SLEEPVol. 49(Supplement_1)
Sleep Research Society (US), Geisinger Neuroscience Institute (US), Geisinger Health System (US)
Good health and well-being
Openalex Percentile: Top 8%
Obstructive Sleep Apnea Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.