Improved coronary CT angiography image quality using photon-counting detector CT in SCAPIS reexamination: scan protocol and comparative analysis

Objectives:The Swedish CArdioPulmonary bioImage Study reexamination (SCAPIS reexamination) is the first population-based study to employ coronary CT angiography (CCTA) using photon-counting detector CT (PCD-CT).It includes 15,000 participants from SCAPIS baseline.This work aims to describe the PCD-CT protocol in SCAPIS reexamination, compare diagnostic image quality with energy-integrating detector CT (EID-CT) at SCAPIS baseline, and assess the comparability of Agatston scores and degree of stenosis between the studies. Methods:The PCD-CT protocol in SCAPIS reexamination is provided.CCTA data from 1,147 participants in SCAPIS reexamination (51% women, age 65 [IQR 61-69]) and 29,554 participants in SCAPIS (52% women, age 58 [54-61]) were analyzed.The image quality in eleven proximal or middle coronary segments was compared, stratified by Agatston score.Agatston scores and stenosis degree were compared in age-and sex-matched samples. Results:Full diagnostic image quality was more frequent in SCAPIS reexamination (1,036/1,147, 90%) compared with SCAPIS baseline (20,468/26,188, 78%), p<0.001, despite a higher calcium burden (Agatston score 8 [0-106] vs. 0 [0-20]).Of participants with Agatston score >400, 89/116, 77%, and 246/945, 26%, had full diagnostic image quality, respectively, p<0.001.Agatston score distributions and stenosis 50% were similar in matched samples (p=0.51 and p=0.08).Of those with 50% stenosis at SCAPIS baseline, 30% (12/40) were reclassified to <50% in SCAPIS reexamination. Conclusion:The optimized PCD-CT protocol in SCAPIS reexamination provided high image quality, irrespective of Agatston scores, outperforming the EID-CT protocol used at SCAPIS baseline.Agatston scores were comparable, but potential differences in stenosis grading warrant further investigation.

Authors

Institutions

Publication Details

Journal
European Journal of Radiology
Published
2026-05-01
DOI
https://doi.org/10.1016/j.ejrad.2026.112920
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improved coronary CT angiography image quality using photon-counting detector CT in SCAPIS reexamination: scan protocol and comparative analysis

Franciska Wikner, Marit Johannesson, Patrik Nowik, Marie‐Louise Aurumskjöld et al.
European Journal of Radiology
Advanced X-ray and CT Imaging
article

Improved coronary CT angiography image quality using photon-counting detector CT in SCAPIS reexamination: scan protocol and comparative analysis

Franciska Wikner, Marit Johannesson, Patrik Nowik, Marie‐Louise Aurumskjöld, Adrian Pistea, Erika Fagman, Andreï Malinovschi, Mats Lidén, Tomas Jernberg, Lilian Henriksson, Gusten Nyberg, Tomasz Baron, Mårten Sandstedt, Monica Segelsjö, Göran Bergström, Henrik Löfmark, David Lind, Anders Persson
article en

Abstract

Objectives:The Swedish CArdioPulmonary bioImage Study reexamination (SCAPIS reexamination) is the first population-based study to employ coronary CT angiography (CCTA) using photon-counting detector CT (PCD-CT).It includes 15,000 participants from SCAPIS baseline.This work aims to describe the PCD-CT protocol in SCAPIS reexamination, compare diagnostic image quality with energy-integrating detector CT (EID-CT) at SCAPIS baseline, and assess the comparability of Agatston scores and degree of stenosis between the studies. Methods:The PCD-CT protocol in SCAPIS reexamination is provided.CCTA data from 1,147 participants in SCAPIS reexamination (51% women, age 65 [IQR 61-69]) and 29,554 participants in SCAPIS (52% women, age 58 [54-61]) were analyzed.The image quality in eleven proximal or middle coronary segments was compared, stratified by Agatston score.Agatston scores and stenosis degree were compared in age-and sex-matched samples. Results:Full diagnostic image quality was more frequent in SCAPIS reexamination (1,036/1,147, 90%) compared with SCAPIS baseline (20,468/26,188, 78%), p<0.001, despite a higher calcium burden (Agatston score 8 [0-106] vs. 0 [0-20]).Of participants with Agatston score >400, 89/116, 77%, and 246/945, 26%, had full diagnostic image quality, respectively, p<0.001.Agatston score distributions and stenosis 50% were similar in matched samples (p=0.51 and p=0.08).Of those with 50% stenosis at SCAPIS baseline, 30% (12/40) were reclassified to <50% in SCAPIS reexamination. Conclusion:The optimized PCD-CT protocol in SCAPIS reexamination provided high image quality, irrespective of Agatston scores, outperforming the EID-CT protocol used at SCAPIS baseline.Agatston scores were comparable, but potential differences in stenosis grading warrant further investigation.

European Journal of Radiology
Linköping University (SE), Uppsala University (SE), Karolinska University Hospital (SE), Uppsala University Hospital (SE), Lund University (SE), Örebro University (SE), Sahlgrenska University Hospital (SE), Karolinska Institutet (SE), Region Västra Götaland (SE), University Hospital of Umeå (SE), Siemens (Sweden) (SE), Skåne University Hospital (SE), Linköping University Hospital (SE), University of Gothenburg (SE), Umeå University (SE)
Hjärt-Lungfonden, Region Örebro län
Openalex Percentile: Top 10%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.