Is the transverse screw really necessary? Biomechanical comparison of fixation strategies in periasetabular osteotomy

Abstract This study aims to compare the biomechanical stability of different screw and plate configurations used in the fixation of acetabular fragments after periacetabular osteotomy (PAO). A total of 42 synthetic hemipelvis models were categorized into six fixation groups: three iliac screws (3 V), four iliac screws (4 V), three screws + lateral compression (LC) screw (3 V + LC), locking three-screw plate (3P), plate + LC screw (3P + LC), and eight-screw plate (4P). All samples were evaluated for stiffness under a constant load of 700 N and for stiffness under maximum load. The greatest failure load was observed in the 3 V group (1249.4 N), whereas the 3P group demonstrated the lowest (658.5 N). In terms of maximum force, both the 3 V and 4 V groups were significantly superior to the 3P (P = .003 and P = .018, respectively) and 3P + LC (P = .006 and P = .021, respectively) groups. Although the 4P configuration showed the highest fracture stiffness (155.1 N/mm), this did not differ significantly among groups. Stiffness at a constant load of 700 N, the 3 V and 3 V + LC groups demonstrated higher stiffness than the 3P and 3P + LC groups, with comparisons to the 3P group approaching statistical significance (P = .064 and P = .063, respectively). Traditional iliac screw structures (especially 3 V and 4 V) provide effective fixation in terms of load-bearing and stiffness, while plate-based systems require a high number of screws and optimal placement to provide adequate stability. Although the LC screw increases stiffness, it provides only a limited contribution to breaking strength.

Authors

Institutions

Publication Details

Journal
Journal of Hip Preservation Surgery
Published
2026-04-14
DOI
https://doi.org/10.1093/jhps/hnag024
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Is the transverse screw really necessary? Biomechanical comparison of fixation strategies in periasetabular osteotomy

Onur Hapa, Eren Akın, Ümit Ali Karabağ, Fatih Ertem et al.
Journal of Hip Preservation Surgery
Orthopaedic implants and arthroplasty
article

Is the transverse screw really necessary? Biomechanical comparison of fixation strategies in periasetabular osteotomy

Onur Hapa, Eren Akın, Ümit Ali Karabağ, Fatih Ertem, Mustafa Celtik, Cihangir Turemis, Selahaddin Aydemir
article en

Abstract

Abstract This study aims to compare the biomechanical stability of different screw and plate configurations used in the fixation of acetabular fragments after periacetabular osteotomy (PAO). A total of 42 synthetic hemipelvis models were categorized into six fixation groups: three iliac screws (3 V), four iliac screws (4 V), three screws + lateral compression (LC) screw (3 V + LC), locking three-screw plate (3P), plate + LC screw (3P + LC), and eight-screw plate (4P). All samples were evaluated for stiffness under a constant load of 700 N and for stiffness under maximum load. The greatest failure load was observed in the 3 V group (1249.4 N), whereas the 3P group demonstrated the lowest (658.5 N). In terms of maximum force, both the 3 V and 4 V groups were significantly superior to the 3P (P = .003 and P = .018, respectively) and 3P + LC (P = .006 and P = .021, respectively) groups. Although the 4P configuration showed the highest fracture stiffness (155.1 N/mm), this did not differ significantly among groups. Stiffness at a constant load of 700 N, the 3 V and 3 V + LC groups demonstrated higher stiffness than the 3P and 3P + LC groups, with comparisons to the 3P group approaching statistical significance (P = .064 and P = .063, respectively). Traditional iliac screw structures (especially 3 V and 4 V) provide effective fixation in terms of load-bearing and stiffness, while plate-based systems require a high number of screws and optimal placement to provide adequate stability. Although the LC screw increases stiffness, it provides only a limited contribution to breaking strength.

Journal of Hip Preservation Surgery
Dokuz Eylül University (TR), Termez State University (UZ), Mustafa Kemal University (TR), Kastamonu University (TR)
Openalex Percentile: Top 6%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.