Closing the evidence gap: Capillary high-sensitivity troponin I verification in an African population using point of care

BACKGROUND: High-sensitivity cardiac troponin (hs-cTnI) assays are central to the early diagnosis and rule-out of acute myocardial infarction (AMI). In resource-limited settings, point-of-care (POC) platforms using capillary blood may reduce turnaround time and improve access to timely decision-making. This study evaluated the performance of capillary hs-cTnI POC sampling in an African population. METHODS: This cross-sectional method comparison study evaluated paired capillary hs-cTnI measurements obtained on POC against corresponding venous serum and plasma samples analysed on a central laboratory assay. Samples were collected from adults presenting with suspected AMI to Charlotte Maxeke Johannesburg Academic Hospital Emergency Department. Analytical verification was performed in accordance with Clinical and Laboratory Standards Institute protocols. Agreement was assessed using Bland Altman analysis. Diagnostic concordance at sex-specific 99th percentile decision thresholds and misclassification rates were determined. RESULTS: = 0.791; 95% CI 0.699-0.857). Mean bias of capillary versus plasma was -15.9% (-173% to +143%) and - 12.9% (-171% to +145%) compared with serum. Serum comparison met the predefined allowable minimum bias of 13.6%. Overall agreement at the 99th percentile was 92.1% (83/91), with substantial inter-method agreement (κ >0.75). No significant asymmetry in discordant classifications was observed (p = 0.7266; 95% CI -8.27 to 3.88). CONCLUSION: Capillary whole-blood hs-cTnI testing demonstrated acceptable analytical agreement and high clinical concordance with central laboratory measurement, supporting its implementation to enhance rapid AMI assessment in resource-limited emergency settings.

Authors

Institutions

Publication Details

Journal
Clinica Chimica Acta
Published
2026-05-01
DOI
https://doi.org/10.1016/j.cca.2026.121054
Primary Topic
Acute Myocardial Infarction Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Closing the evidence gap: Capillary high-sensitivity troponin I verification in an African population using point of care

Khoza Siyabonga, Maseti Amvuyele Ann, Chikomba Chemedzai, Jacob Doreen
Clinica Chimica Acta
Acute Myocardial Infarction Research
article

Closing the evidence gap: Capillary high-sensitivity troponin I verification in an African population using point of care

Khoza Siyabonga, Maseti Amvuyele Ann, Chikomba Chemedzai, Jacob Doreen
article en

Abstract

BACKGROUND: High-sensitivity cardiac troponin (hs-cTnI) assays are central to the early diagnosis and rule-out of acute myocardial infarction (AMI). In resource-limited settings, point-of-care (POC) platforms using capillary blood may reduce turnaround time and improve access to timely decision-making. This study evaluated the performance of capillary hs-cTnI POC sampling in an African population. METHODS: This cross-sectional method comparison study evaluated paired capillary hs-cTnI measurements obtained on POC against corresponding venous serum and plasma samples analysed on a central laboratory assay. Samples were collected from adults presenting with suspected AMI to Charlotte Maxeke Johannesburg Academic Hospital Emergency Department. Analytical verification was performed in accordance with Clinical and Laboratory Standards Institute protocols. Agreement was assessed using Bland Altman analysis. Diagnostic concordance at sex-specific 99th percentile decision thresholds and misclassification rates were determined. RESULTS: = 0.791; 95% CI 0.699-0.857). Mean bias of capillary versus plasma was -15.9% (-173% to +143%) and - 12.9% (-171% to +145%) compared with serum. Serum comparison met the predefined allowable minimum bias of 13.6%. Overall agreement at the 99th percentile was 92.1% (83/91), with substantial inter-method agreement (κ >0.75). No significant asymmetry in discordant classifications was observed (p = 0.7266; 95% CI -8.27 to 3.88). CONCLUSION: Capillary whole-blood hs-cTnI testing demonstrated acceptable analytical agreement and high clinical concordance with central laboratory measurement, supporting its implementation to enhance rapid AMI assessment in resource-limited emergency settings.

Clinica Chimica Acta
National Health Laboratory Service (ZA)
Siemens Medical Solutions USA
Zero hunger
Openalex Percentile: Top 8%
Acute Myocardial Infarction Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Closing the evidence gap: Capillary high-sensitivity troponin I verification in an African population using point of care — Khoza Siyabonga, Maseti Amvuyele Ann, et al. · Clinica Chimica Acta (2026) | TGRS Research Map | TGRS