Assessing the technical capability of a room- and a gantry-mounted kV imaging device for intra-fractional fluoroscopy during stereotactic lung radiotherapy at a C-arm linear accelerator
Background and Purpose: Monitoring lung tumors during stereotactic radiotherapy is important for managing breathing motion, yet the high-performance imaging modules on C-arm linear accelerators are mainly used for patient positioning. This study benchmarked a room-mounted against a gantry-mounted kilovolt (kV) imaging device to assess their potential for lung tumor fluoroscopy. Materials and Methods: The investigated systems were the ExacTrac Dynamic (ETD, Brainlab, Germany) and the X-ray Volumetric Imaging (XVI, Elekta, Sweden). Contrast-to-Noise Ratio (CNR) and entrance air kerma were measured for different imaging settings - tube voltage (70 kV to 130 kV), current (10 mA to 320 mA), exposure time (100 ms and 40 ms) - using two anthropomorphic thorax phantoms housing three tumors. Additionally, CNR change as a function of time, achievable fluoroscopy imaging duration based on X-ray generator and anode heat and X-ray tube cool-down were evaluated. Results: CNRs agreed within 15% for the two systems and changed less than 2% over the investigated fluoroscopy durations. Using default imaging settings and considering the average clinical duration of stereotactic lung treatments, entrance air kermas were 44 mGy (room-mounted) and 41 mGy (gantry-mounted). Achievable fluoroscopy durations with default settings were 87 s and 1350 s, respectively. Anode cool-down times were 125 min (room-mounted) and 215 min (gantry-mounted). Conclusion: The performance metrics of the room-mounted imaging system aligned with those of the gantry-mounted device with regard to CNR and air kerma. The room-mounted system showed a preferable cool-down behavior. However, adaptation of imaging settings is necessary to increase the fluoroscopy duration and avoid overheating.
Authors
- Martin Buschmann (ORCID: https://orcid.org/0000-0001-9946-1939)
- Dietmar Georg (ORCID: https://orcid.org/0000-0002-8327-3877)
- Barbara Knäusl (ORCID: https://orcid.org/0000-0003-4031-1711)
- Hannah Jungreuthmayer
- M. Schmid (ORCID: https://orcid.org/0000-0003-4061-3605)
- Andreas Renner (ORCID: https://orcid.org/0000-0001-9813-0919)
- Julius Arnold (ORCID: https://orcid.org/0009-0003-5551-5423)
- Wolfgang Lechner (ORCID: https://orcid.org/0000-0001-9211-7510)
Institutions
- Christian Doppler Laboratory for Thermoelectricity (AT)
- Medical University of Vienna (AT)
Publication Details
- Journal
- Physics and Imaging in Radiation Oncology
- Published
- 2026-05-01
- DOI
- https://doi.org/10.1016/j.phro.2026.100988
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Christian Doppler Forschungsgesellschaft