Assessing the technical capability of a room- and a gantry-mounted kV imaging device for intra-fractional fluoroscopy during stereotactic lung radiotherapy at a C-arm linear accelerator

Background and Purpose: Monitoring lung tumors during stereotactic radiotherapy is important for managing breathing motion, yet the high-performance imaging modules on C-arm linear accelerators are mainly used for patient positioning. This study benchmarked a room-mounted against a gantry-mounted kilovolt (kV) imaging device to assess their potential for lung tumor fluoroscopy. Materials and Methods: The investigated systems were the ExacTrac Dynamic (ETD, Brainlab, Germany) and the X-ray Volumetric Imaging (XVI, Elekta, Sweden). Contrast-to-Noise Ratio (CNR) and entrance air kerma were measured for different imaging settings - tube voltage (70 kV to 130 kV), current (10 mA to 320 mA), exposure time (100 ms and 40 ms) - using two anthropomorphic thorax phantoms housing three tumors. Additionally, CNR change as a function of time, achievable fluoroscopy imaging duration based on X-ray generator and anode heat and X-ray tube cool-down were evaluated. Results: CNRs agreed within 15% for the two systems and changed less than 2% over the investigated fluoroscopy durations. Using default imaging settings and considering the average clinical duration of stereotactic lung treatments, entrance air kermas were 44 mGy (room-mounted) and 41 mGy (gantry-mounted). Achievable fluoroscopy durations with default settings were 87 s and 1350 s, respectively. Anode cool-down times were 125 min (room-mounted) and 215 min (gantry-mounted). Conclusion: The performance metrics of the room-mounted imaging system aligned with those of the gantry-mounted device with regard to CNR and air kerma. The room-mounted system showed a preferable cool-down behavior. However, adaptation of imaging settings is necessary to increase the fluoroscopy duration and avoid overheating.

Authors

Institutions

Publication Details

Journal
Physics and Imaging in Radiation Oncology
Published
2026-05-01
DOI
https://doi.org/10.1016/j.phro.2026.100988
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing the technical capability of a room- and a gantry-mounted kV imaging device for intra-fractional fluoroscopy during stereotactic lung radiotherapy at a C-arm linear accelerator

Martin Buschmann, Dietmar Georg, Barbara Knäusl, Hannah Jungreuthmayer et al.
Physics and Imaging in Radiation Oncology
Advanced Radiotherapy Techniques
article

Assessing the technical capability of a room- and a gantry-mounted kV imaging device for intra-fractional fluoroscopy during stereotactic lung radiotherapy at a C-arm linear accelerator

Martin Buschmann, Dietmar Georg, Barbara Knäusl, Hannah Jungreuthmayer, M. Schmid, Andreas Renner, Julius Arnold, Wolfgang Lechner
article en

Abstract

Background and Purpose: Monitoring lung tumors during stereotactic radiotherapy is important for managing breathing motion, yet the high-performance imaging modules on C-arm linear accelerators are mainly used for patient positioning. This study benchmarked a room-mounted against a gantry-mounted kilovolt (kV) imaging device to assess their potential for lung tumor fluoroscopy. Materials and Methods: The investigated systems were the ExacTrac Dynamic (ETD, Brainlab, Germany) and the X-ray Volumetric Imaging (XVI, Elekta, Sweden). Contrast-to-Noise Ratio (CNR) and entrance air kerma were measured for different imaging settings - tube voltage (70 kV to 130 kV), current (10 mA to 320 mA), exposure time (100 ms and 40 ms) - using two anthropomorphic thorax phantoms housing three tumors. Additionally, CNR change as a function of time, achievable fluoroscopy imaging duration based on X-ray generator and anode heat and X-ray tube cool-down were evaluated. Results: CNRs agreed within 15% for the two systems and changed less than 2% over the investigated fluoroscopy durations. Using default imaging settings and considering the average clinical duration of stereotactic lung treatments, entrance air kermas were 44 mGy (room-mounted) and 41 mGy (gantry-mounted). Achievable fluoroscopy durations with default settings were 87 s and 1350 s, respectively. Anode cool-down times were 125 min (room-mounted) and 215 min (gantry-mounted). Conclusion: The performance metrics of the room-mounted imaging system aligned with those of the gantry-mounted device with regard to CNR and air kerma. The room-mounted system showed a preferable cool-down behavior. However, adaptation of imaging settings is necessary to increase the fluoroscopy duration and avoid overheating.

Physics and Imaging in Radiation OncologyVol. 39
Christian Doppler Laboratory for Thermoelectricity (AT), Medical University of Vienna (AT)
Christian Doppler Forschungsgesellschaft
Openalex Percentile: Top 9%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.