High-Resolution 2D versus 3D Lumbar Spine MRI Optimized with a Deep Learning Reconstruction Algorithm and Prototype Conformal Coil

BACKGROUND AND PURPOSE: Lumbar spine MRI is predominantly performed using 2D FSE sequences. 3D FSE sequences offer potential advantages over 2D, especially with recent advances in deep-learning reconstruction (DLR) and the use of conformal high-density coil arrays to improve the SNR. This study aimed to compare image quality and diagnostically relevant differences between 2D and 3D lumbar spine protocols, both using a prototype conformal coil and DLR. MATERIALS AND METHODS: =31). A prototype, conformal spine coil array with 70 total posterior receiver elements was used at 3T. DLR was applied to both the 2D- and 3D-protocols to improve sharpness and reduce noise. For 3D, a prototype DLR ("Sonic DL") technique was applied to provide higher acceleration factors. Three attending musculoskeletal radiologists assessed 2D and 3D T1-weighted (T1), T2-weighted (T2), and STIR sequences for image quality and diagnostic performance. The Gwet agreement coefficient (AC1, AC2) or intraclass correlation (ICC) was used to assess interrater agreement, while the Wilcoxon signed rank test or a sign test with Bonferroni correction assessed differences between 2D and 3D sequences. RESULTS: There was an overall 14% scan time reduction of 3D compared with 2D sequences, and a 38% time reduction for T2 sequences. All 3D sequences exhibited similar image quality to 2D. 3D T2 sequences had less CSF flow artifact than 2D T2 sequences. Interrater agreement was mostly comparable between all 2D and 3D assessments, including excellent agreement for the spinal canal area (ICC = 0.952 [95% CI: 0.907-0.976] for 2D, 0.941 [0.866-0.973] for 3D), good agreement for foraminal stenosis assessments on T2 sequences (AC2 = 0.713 to 0.795 for 2D, 0.706-0.832 for 3D), and good agreement for fracture detection on T1 sequences (AC1 = 0.869 for 2D, 0.815 for 3D). CONCLUSIONS: DLR-enabled 3D lumbar spine MRI offers comparable image quality to 2D with reduced overall scan time.

Authors

Institutions

Publication Details

Journal
American Journal of Neuroradiology
Published
2026-05-01
DOI
https://doi.org/10.3174/ajnr.a9141
Primary Topic
Medical Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-Resolution 2D versus 3D Lumbar Spine MRI Optimized with a Deep Learning Reconstruction Algorithm and Prototype Conformal Coil

Yun-Jeong Stickle, Mark Giancola, Daniel M. Alschuler, Allison Lowe et al.
American Journal of Neuroradiology
Medical Imaging and Analysis
article

High-Resolution 2D versus 3D Lumbar Spine MRI Optimized with a Deep Learning Reconstruction Algorithm and Prototype Conformal Coil

Yun-Jeong Stickle, Mark Giancola, Daniel M. Alschuler, Allison Lowe, Richard Kijowski, Shu-Han Wang, Ek T. Tan, Meghan Sahr, Clyve Follante, J. Levi Chazen, Fraser Robb, Darryl B. Sneag, Yan Wen
article en

Abstract

BACKGROUND AND PURPOSE: Lumbar spine MRI is predominantly performed using 2D FSE sequences. 3D FSE sequences offer potential advantages over 2D, especially with recent advances in deep-learning reconstruction (DLR) and the use of conformal high-density coil arrays to improve the SNR. This study aimed to compare image quality and diagnostically relevant differences between 2D and 3D lumbar spine protocols, both using a prototype conformal coil and DLR. MATERIALS AND METHODS: =31). A prototype, conformal spine coil array with 70 total posterior receiver elements was used at 3T. DLR was applied to both the 2D- and 3D-protocols to improve sharpness and reduce noise. For 3D, a prototype DLR ("Sonic DL") technique was applied to provide higher acceleration factors. Three attending musculoskeletal radiologists assessed 2D and 3D T1-weighted (T1), T2-weighted (T2), and STIR sequences for image quality and diagnostic performance. The Gwet agreement coefficient (AC1, AC2) or intraclass correlation (ICC) was used to assess interrater agreement, while the Wilcoxon signed rank test or a sign test with Bonferroni correction assessed differences between 2D and 3D sequences. RESULTS: There was an overall 14% scan time reduction of 3D compared with 2D sequences, and a 38% time reduction for T2 sequences. All 3D sequences exhibited similar image quality to 2D. 3D T2 sequences had less CSF flow artifact than 2D T2 sequences. Interrater agreement was mostly comparable between all 2D and 3D assessments, including excellent agreement for the spinal canal area (ICC = 0.952 [95% CI: 0.907-0.976] for 2D, 0.941 [0.866-0.973] for 3D), good agreement for foraminal stenosis assessments on T2 sequences (AC2 = 0.713 to 0.795 for 2D, 0.706-0.832 for 3D), and good agreement for fracture detection on T1 sequences (AC1 = 0.869 for 2D, 0.815 for 3D). CONCLUSIONS: DLR-enabled 3D lumbar spine MRI offers comparable image quality to 2D with reduced overall scan time.

American Journal of NeuroradiologyVol. 47(5)
Hospital for Special Surgery (US), General Electric (Spain) (ES)
GE Healthcare
Openalex Percentile: Top 11%
Medical Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.