Biomechanical Interaction of Additively Manufactured Ti6Al4V Metamaterial Inserts with Vascular Tissue: Effects of Surface Topography on Tissue Damage and Residual Stress Distribution

This study evaluates the biomechanical response of vascular tissue subjected to compression by mechanical metamaterial-based inserts fabricated from Ti6Al4V alloy using additive manufacturing. The inserts featured varying surface topographies: smooth, moderately notched, and deeply notched; and internal void geometries: circular, elliptical, triangular. Ring-shaped porcine aorta samples were compressed under controlled conditions for 10 minutes in a simulated physiological environment. Digital microscopy and histological analysis revealed tissue delamination and a reduction in vessel wall thickness, with delamination observed in up to 90% of the reference samples and wall thickness ranging from ∼1.2 mm for the reference to ∼2.4–3.0 mm for the modified configurations. While tissue alterations were observed, no visible deformation of the metallic inserts was detected post-testing, confirming their structural integrity and the precision of the 3D printing process. Therefore, the material's response to pressure on a blood vessel was examined in terms of structural deformation, i.e., changes in the interplanar distance manifested in the form of residual stresses. The stress distribution was measured by X-ray diffraction tomography. X-ray diffraction measurements indicated the presence of high-magnitude compressive residual stresses, especially in samples with deeper surface structuring. These compressive stresses-confined within individual grains – are associated with enhanced resistance to crack initiation and improved fatigue performance. Finite element modeling supported experimental observations, showing increased stress with greater surface complexity. Overall, the results demonstrate that surface morphology in mechanical metamaterial inserts critically influences both tissue interaction and residual stress distribution, offering guidance for the design of durable and tissue-safe surgical tools. • Surface topography of Ti6Al4V inserts governs stress transfer to vascular tissue • Deep surface structuring increases tissue damage and compressive residual stresses • Additively manufactured metamaterials show high structural stability under loading • XRD reveals depth-dependent compressive residual stresses after vessel compression • Moderate surface structuring reduces vascular trauma while preserving durability

Authors

Institutions

Publication Details

Journal
Journal of Materials Research and Technology
Published
2026-05-01
DOI
https://doi.org/10.1016/j.jmrt.2026.05.019
Primary Topic
Cellular and Composite Structures
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical Interaction of Additively Manufactured Ti6Al4V Metamaterial Inserts with Vascular Tissue: Effects of Surface Topography on Tissue Damage and Residual Stress Distribution

Roman Major, Marcin Dyner, Łukasz Major, Aneta Dyner et al.
Journal of Materials Research and Technology
Cellular and Composite Structures
article

Biomechanical Interaction of Additively Manufactured Ti6Al4V Metamaterial Inserts with Vascular Tissue: Effects of Surface Topography on Tissue Damage and Residual Stress Distribution

Roman Major, Marcin Dyner, Łukasz Major, Aneta Dyner, Zuzanna Zajac, Enbiya Türedi, Karolina Szawiraacz, Janusz Szewczenko, Magdalena Kopernik, Przemysław Kurtyka, Ewa Jasek-Gajda, Serap Gümüş, Małgorzata Pomorska, Jurgen M. Lackner, Marcin Basiaga, Christian Pfeifer, Justyna Więcek-Chmielarz
article en

Abstract

This study evaluates the biomechanical response of vascular tissue subjected to compression by mechanical metamaterial-based inserts fabricated from Ti6Al4V alloy using additive manufacturing. The inserts featured varying surface topographies: smooth, moderately notched, and deeply notched; and internal void geometries: circular, elliptical, triangular. Ring-shaped porcine aorta samples were compressed under controlled conditions for 10 minutes in a simulated physiological environment. Digital microscopy and histological analysis revealed tissue delamination and a reduction in vessel wall thickness, with delamination observed in up to 90% of the reference samples and wall thickness ranging from ∼1.2 mm for the reference to ∼2.4–3.0 mm for the modified configurations. While tissue alterations were observed, no visible deformation of the metallic inserts was detected post-testing, confirming their structural integrity and the precision of the 3D printing process. Therefore, the material's response to pressure on a blood vessel was examined in terms of structural deformation, i.e., changes in the interplanar distance manifested in the form of residual stresses. The stress distribution was measured by X-ray diffraction tomography. X-ray diffraction measurements indicated the presence of high-magnitude compressive residual stresses, especially in samples with deeper surface structuring. These compressive stresses-confined within individual grains – are associated with enhanced resistance to crack initiation and improved fatigue performance. Finite element modeling supported experimental observations, showing increased stress with greater surface complexity. Overall, the results demonstrate that surface morphology in mechanical metamaterial inserts critically influences both tissue interaction and residual stress distribution, offering guidance for the design of durable and tissue-safe surgical tools. • Surface topography of Ti6Al4V inserts governs stress transfer to vascular tissue • Deep surface structuring increases tissue damage and compressive residual stresses • Additively manufactured metamaterials show high structural stability under loading • XRD reveals depth-dependent compressive residual stresses after vessel compression • Moderate surface structuring reduces vascular trauma while preserving durability

Journal of Materials Research and Technology
Silesian University of Technology (PL), Jan Długosz University (PL), Jagiellonian University (PL), Joanneum Research (AT), Mari State University (RU), Institute of Metallurgy and Materials Science (PL), AustriaTech (Austria) (AT), Foundation of Cardiac Surgery Development (PL), Kocaeli Üniversitesi (TR), AGH University of Krakow (PL), Polish Academy of Sciences (PL)
Openalex Percentile: Top 11%
Cellular and Composite Structures
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.