Mechanical Fatigue of Titanium Dental Implants After Implantoplasty: An In Vitro Study Combined with Finite Element Simulations

The increasing prevalence of peri-implantitis has led to a growing clinical use of implantoplasty, a procedure involving intraoral machining of the dental implant surface to remove biofilm. The absence of standardized clinical protocols may contribute to premature fatigue failure of dental implants. The present study aimed to evaluate the influence of machining depth on the cyclic mechanical behavior of dental implants. A total of 250 commercially pure grade 4 titanium dental implants were distributed into four groups according to machining depth: untreated (original), 0.2 mm, 0.4 mm, and 0.6 mm wall reduction. The implant system featured an internal connection with a thread height of 0.4 mm. Finite element analysis was performed for each machining depth to evaluate von Mises stress distribution and simulate fatigue behavior. The numerical models were validated through experimental fatigue testing using a servo-hydraulic MTS Bionix testing machine under ISO 14801:2016 conditions, showing a high correlation between simulated and experimental results (correlation coefficients > 0.9). The results indicated that maximum von Mises stresses were concentrated at the junction between the implant thread and the implant body. The fatigue limit of the untreated implants was approximately 351 N. Implants subjected to 0.4 mm machining exhibited a fatigue limit of 301 N, whereas lower fatigue limits were observed for 0.2 mm (255 N) and 0.6 mm (185 N) reductions. These findings suggest a significant mechanical effect of thread removal: 0.4 mm implantoplasty may provide improved fatigue performance compared to 0.2 mm, potentially due to reduced stress concentration at the thread–body junction. At high applied loads, fracture occurred in the coronal region of the implant, whereas at lower loads failure shifted to the implant–abutment connection. Although a good agreement between numerical and experimental results was observed, these findings should be interpreted with caution due to the in vitro testing conditions and the assumptions inherent to the finite element simulations. Therefore, while the results suggest that implantoplasty depth should not exceed the original thread height, further validation under clinically relevant conditions is required to confirm its impact on long-term mechanical reliability.

Authors

Institutions

Publication Details

Journal
Journal of Functional Biomaterials
Published
2026-05-02
DOI
https://doi.org/10.3390/jfb17050221
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanical Fatigue of Titanium Dental Implants After Implantoplasty: An In Vitro Study Combined with Finite Element Simulations

Pablo Sevilla, M. Cerrolaza, Esteban Padullés-Roig, Eugenio Velasco‐Ortega et al.
Journal of Functional Biomaterials
Dental Implant Techniques and Outcomes
article

Mechanical Fatigue of Titanium Dental Implants After Implantoplasty: An In Vitro Study Combined with Finite Element Simulations

Pablo Sevilla, M. Cerrolaza, Esteban Padullés-Roig, Eugenio Velasco‐Ortega, Javier Gil, Jeanne Parache, Darcio Fonseca, Conrado Aparicio
article en

Abstract

The increasing prevalence of peri-implantitis has led to a growing clinical use of implantoplasty, a procedure involving intraoral machining of the dental implant surface to remove biofilm. The absence of standardized clinical protocols may contribute to premature fatigue failure of dental implants. The present study aimed to evaluate the influence of machining depth on the cyclic mechanical behavior of dental implants. A total of 250 commercially pure grade 4 titanium dental implants were distributed into four groups according to machining depth: untreated (original), 0.2 mm, 0.4 mm, and 0.6 mm wall reduction. The implant system featured an internal connection with a thread height of 0.4 mm. Finite element analysis was performed for each machining depth to evaluate von Mises stress distribution and simulate fatigue behavior. The numerical models were validated through experimental fatigue testing using a servo-hydraulic MTS Bionix testing machine under ISO 14801:2016 conditions, showing a high correlation between simulated and experimental results (correlation coefficients > 0.9). The results indicated that maximum von Mises stresses were concentrated at the junction between the implant thread and the implant body. The fatigue limit of the untreated implants was approximately 351 N. Implants subjected to 0.4 mm machining exhibited a fatigue limit of 301 N, whereas lower fatigue limits were observed for 0.2 mm (255 N) and 0.6 mm (185 N) reductions. These findings suggest a significant mechanical effect of thread removal: 0.4 mm implantoplasty may provide improved fatigue performance compared to 0.2 mm, potentially due to reduced stress concentration at the thread–body junction. At high applied loads, fracture occurred in the coronal region of the implant, whereas at lower loads failure shifted to the implant–abutment connection. Although a good agreement between numerical and experimental results was observed, these findings should be interpreted with caution due to the in vitro testing conditions and the assumptions inherent to the finite element simulations. Therefore, while the results suggest that implantoplasty depth should not exceed the original thread height, further validation under clinically relevant conditions is required to confirm its impact on long-term mechanical reliability.

Journal of Functional BiomaterialsVol. 17(5)
Institució Catalana de Recerca i Estudis Avançats (ES), Valencian International University (ES), Universitat Internacional de Catalunya (ES), Universitat Ramon Llull (ES), Universidad de Sevilla (ES), Universitat Politècnica de Catalunya (ES)
Agencia Estatal de Investigación
Openalex Percentile: Top 7%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.