Accurate monitoring of respiratory rates is critical for early detection of a range of clinical conditions. However, standard manual counting or inadequate clinical monitoring often fails to provide reliable measurements. This study evaluated and validated the PneumoWave biosensor for respiratory rate measurement across a broad physiological range and different body postures (45°, 90°, and 180°) in both in vitro and in vivo settings. In vitro validation was performed using a SimMan ALS manikin operated at respiratory settings of 6–30 breaths per minute, with 10 s periods of simulated apnea. In vivo validation involved 20 healthy volunteers performing metronome-guided breathing while wearing bilateral PneumoWave biosensors. In vitro results demonstrated an excellent correlation between biosensors and manikin respiratory settings and captured all apnea events (r = 0.99, ICC = 0.99). In vivo findings showed good agreement with direct observational count (r = 0.99, R2 = 0.99, ICC = 0.99), with 97% of apnea events captured by both devices in all positions. Body postures had no significant impact on biosensor accuracy. These findings demonstrate that the PneumoWave biosensor provides accurate and reliable respiratory monitoring and supports its potential as a robust, non-invasive tool for continuous clinical and remote patient monitoring.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2026-05-01
DOI
https://doi.org/10.3390/bios16050256
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Respiratory Rate and Simulated Apnea Utilizing the PneumoWave Biosensor: In Vitro and In Vivo Validation

Burcu Kolukisa Birgec, Alexander Balfour Mullen, Beyza Toprak
Biosensors
Non-Invasive Vital Sign Monitoring
article

Assessment of Respiratory Rate and Simulated Apnea Utilizing the PneumoWave Biosensor: In Vitro and In Vivo Validation

Burcu Kolukisa Birgec, Alexander Balfour Mullen, Beyza Toprak
article en

Abstract

Accurate monitoring of respiratory rates is critical for early detection of a range of clinical conditions. However, standard manual counting or inadequate clinical monitoring often fails to provide reliable measurements. This study evaluated and validated the PneumoWave biosensor for respiratory rate measurement across a broad physiological range and different body postures (45°, 90°, and 180°) in both in vitro and in vivo settings. In vitro validation was performed using a SimMan ALS manikin operated at respiratory settings of 6–30 breaths per minute, with 10 s periods of simulated apnea. In vivo validation involved 20 healthy volunteers performing metronome-guided breathing while wearing bilateral PneumoWave biosensors. In vitro results demonstrated an excellent correlation between biosensors and manikin respiratory settings and captured all apnea events (r = 0.99, ICC = 0.99). In vivo findings showed good agreement with direct observational count (r = 0.99, R2 = 0.99, ICC = 0.99), with 97% of apnea events captured by both devices in all positions. Body postures had no significant impact on biosensor accuracy. These findings demonstrate that the PneumoWave biosensor provides accurate and reliable respiratory monitoring and supports its potential as a robust, non-invasive tool for continuous clinical and remote patient monitoring.

BiosensorsVol. 16(5)
University of Strathclyde (GB)
Good health and well-being
Openalex Percentile: Top 10%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.