Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT

The ramp filter kernel and cutoff frequency are fundamental parameters of the Feldkamp–Davis–Kress (FDK) algorithm that determine the resolution and noise characteristics of the reconstructed image. Despite their importance, the filter configuration is frequently unreported in preclinical micro-CT studies. We reconstruct identical data from a GE eXplore CT 120 scanner using four filter kernels (ramp, Shepp-Logan, cosine, Hamming) at four cutoff frequencies (1.0, 0.8, 0.6, and 0.379 × Nyquist, matched to the detector-to-voxel size ratio) and evaluate each of the sixteen configurations using the modulation transfer function (MTF), noise power spectrum (NPS), and non-prewhitening detectability index (NPW d ′ ). Qualitative assessment is performed on a mouse lung specimen. Across the sixteen configurations, MTF 10 ranges from 1.74 to 3.22 lp/mm, integrated NPS from 75 670 to 13 259 HU 2 , and the Rose criterion crossing diameter from 2.03 to 0.90 mm at Δ C = 500 H U and from 6.89 to 3.56 mm at 100 HU. This note presents the data as a concise, primarily comparative visual and quantitative reference for FDK filter selection in preclinical cone-beam CT.

Authors

Publication Details

Journal
PLoS ONE
Published
2026-08-18
DOI
https://doi.org/10.1371/journal.pone.0349360
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT

Falk Ludwig Wiegmann, Nancy L Ford
PLoS ONE
Medical Imaging Techniques and Applications
article

Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT

Falk Ludwig Wiegmann, Nancy L Ford
article en

Abstract

The ramp filter kernel and cutoff frequency are fundamental parameters of the Feldkamp–Davis–Kress (FDK) algorithm that determine the resolution and noise characteristics of the reconstructed image. Despite their importance, the filter configuration is frequently unreported in preclinical micro-CT studies. We reconstruct identical data from a GE eXplore CT 120 scanner using four filter kernels (ramp, Shepp-Logan, cosine, Hamming) at four cutoff frequencies (1.0, 0.8, 0.6, and 0.379 × Nyquist, matched to the detector-to-voxel size ratio) and evaluate each of the sixteen configurations using the modulation transfer function (MTF), noise power spectrum (NPS), and non-prewhitening detectability index (NPW d ′ ). Qualitative assessment is performed on a mouse lung specimen. Across the sixteen configurations, MTF 10 ranges from 1.74 to 3.22 lp/mm, integrated NPS from 75 670 to 13 259 HU 2 , and the Rose criterion crossing diameter from 2.03 to 0.90 mm at Δ C = 500 H U and from 6.89 to 3.56 mm at 100 HU. This note presents the data as a concise, primarily comparative visual and quantitative reference for FDK filter selection in preclinical cone-beam CT.

PLoS ONEVol. 21(8)
Openalex Percentile: Top 54%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.