Wirelessly Interrogated, Implantable Capacitive MEMS Sensors for Continuous Intraocular Pressure Monitoring
This work presents wirelessly interrogated microelectromechanical system (MEMS) capacitive sensors for continuous intraocular pressure (IOP) monitoring. The sensor uses a passive inductor–capacitor (LC) tank circuit comprising a fixed, on-chip spiral inductor and a pressure-sensitive, variable-gap capacitor with parallel-plate membrane electrodes and side anchors. The membrane is designed with dimensions of 500 µm × 500 µm × 2 µm and a capacitive transducer gap of 2.5 µm. Applied pressure deflects the top membrane, producing a corresponding capacitance variation that changes the frequency and phase response of the LC tank circuit, enabling real-time and continuous IOP monitoring over a target detection range of 0–50 mmHg and beyond. Mutual inductive coupling between the sensor and the external readout coil is investigated as a reliable readout mechanism.
Authors
- Adnan Zaman (ORCID: https://orcid.org/0000-0003-0351-3146)
- Liguan Li (ORCID: https://orcid.org/0009-0000-1655-4583)
- Ramesh Ayyala
- Jing Wang (ORCID: https://orcid.org/0000-0002-9380-4467)
Institutions
- King Abdulaziz City for Science and Technology (SA)
- University of South Florida (US)
Publication Details
- Journal
- Sensors
- Published
- 2026-04-30
- DOI
- https://doi.org/10.3390/s26092806
- Primary Topic
- Wireless Power Transfer Systems
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- U.S. Army Medical Research and Development Command