Single-sequence low-field 3D T1-weighted MRI for scaphoid fracture detection compared to radiography: prospective diagnostic accuracy pilot study

PURPOSE: To determine the diagnostic accuracy of a novel high-resolution 3D T1-weighted gradient-echo sequence in low-field MRI for detecting scaphoid fractures compared with radiography. METHODS: In this prospective single-center study, patients with suspected scaphoid injury underwent wrist radiography and 0.31 T extremity MRI including a 4-min 3D gradient-echo sequence (3D SST1). Three musculoskeletal radiologists independently evaluated radiographs and the isolated 3D sequence in separate sessions, rating fracture presence, displacement/stability, diagnostic confidence (five-point Likert scale), and need for additional imaging. The reference standard was based on CT and standard MRI sequences, evaluated by two experienced musculoskeletal radiologists. RESULTS: Forty patients with 11 scaphoid fractures were included. Across readers, AUC for scaphoid fracture detection was 1.000 for 3D MRI and 0.823 for radiography (AUC-difference 0.177, 95% CI: -0.004-0.359, p = 0.056). For 3D MRI, sensitivity and NPV were 1.000 for all readers. Specificity (0.931-0.966 vs 0.655-0.862) and PPV (0.846-0.917 vs 0.444-0.667) were higher than for radiography. Diagnostic confidence was excellent (median score 1) with near-perfect inter-rater agreement (κ = 0.961). Additional imaging was not recommended in 40/36/37 cases (reader 1/2/3). CONCLUSION: A single 4-min high-resolution 3D low-field MRI sequence enabled highly accurate detection of scaphoid fractures with perfect sensitivity and NPV, a high reader confidence, and an improved performance compared to radiography in this small pilot cohort. This approach may facilitate rapid, radiation-free screening in acute wrist trauma.

Authors

Institutions

Publication Details

Journal
Emergency Radiology
Published
2026-05-02
DOI
https://doi.org/10.1007/s10140-026-02481-3
Primary Topic
Orthopedic Surgery and Rehabilitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Single-sequence low-field 3D T1-weighted MRI for scaphoid fracture detection compared to radiography: prospective diagnostic accuracy pilot study

Horst Zajonc, Claudia Neubauer, Sebastian M. Goerke, Filip Šimunović et al.
Emergency Radiology
Orthopedic Surgery and Rehabilitation
article

Single-sequence low-field 3D T1-weighted MRI for scaphoid fracture detection compared to radiography: prospective diagnostic accuracy pilot study

Horst Zajonc, Claudia Neubauer, Sebastian M. Goerke, Filip Šimunović, Jakob Neubauer, Christopher Schuppert, Carolin Sophie Reidelbach, Sebastian Wirtz, F Bamberg, M Russe, Thierno D Diallo
article en

Abstract

PURPOSE: To determine the diagnostic accuracy of a novel high-resolution 3D T1-weighted gradient-echo sequence in low-field MRI for detecting scaphoid fractures compared with radiography. METHODS: In this prospective single-center study, patients with suspected scaphoid injury underwent wrist radiography and 0.31 T extremity MRI including a 4-min 3D gradient-echo sequence (3D SST1). Three musculoskeletal radiologists independently evaluated radiographs and the isolated 3D sequence in separate sessions, rating fracture presence, displacement/stability, diagnostic confidence (five-point Likert scale), and need for additional imaging. The reference standard was based on CT and standard MRI sequences, evaluated by two experienced musculoskeletal radiologists. RESULTS: Forty patients with 11 scaphoid fractures were included. Across readers, AUC for scaphoid fracture detection was 1.000 for 3D MRI and 0.823 for radiography (AUC-difference 0.177, 95% CI: -0.004-0.359, p = 0.056). For 3D MRI, sensitivity and NPV were 1.000 for all readers. Specificity (0.931-0.966 vs 0.655-0.862) and PPV (0.846-0.917 vs 0.444-0.667) were higher than for radiography. Diagnostic confidence was excellent (median score 1) with near-perfect inter-rater agreement (κ = 0.961). Additional imaging was not recommended in 40/36/37 cases (reader 1/2/3). CONCLUSION: A single 4-min high-resolution 3D low-field MRI sequence enabled highly accurate detection of scaphoid fractures with perfect sensitivity and NPV, a high reader confidence, and an improved performance compared to radiography in this small pilot cohort. This approach may facilitate rapid, radiation-free screening in acute wrist trauma.

Emergency Radiology
University of Freiburg (DE), University Medical Center Freiburg (DE), Urology Team (US)
Quality Education
Openalex Percentile: Top 12%
Orthopedic Surgery and Rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.