Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts

BACKGROUND: Hemocompatibility-related adverse events still limit the broader adoption of rotodynamic blood pumps. This study aims to evaluate the potential of a multistage pump design to enhance hemocompatibility in left ventricular assist devices. METHODS: Computational fluid dynamic simulations were utilized to compare the pump performance and hemocompatibility of a Two-Stage Pump to two equivalent single-stage devices, designed for identical operating conditions. One single-stage design features equal impeller and volute dimensions; the other, equal specific speed and specific diameter. Our simulation setup was validated against in vitro experiments of the pump performance of the Two-Stage Pump. RESULTS: The Two-Stage Pump achieved a substantial reduction in circumferential velocity (24% and 29% reduction). Although the two-stage configuration thereby effectively reduces shear stress, it simultaneously translates to a longer residence time due to the larger priming volume. However, because the numerical blood damage estimation weighs shear stress exponentially more than residence time, the beneficial effect of reduced shear stress partially compensates for the longer residence time. As a result, the Two-Stage Pump demonstrates a lower hemolysis index (12% and 14% reduction) compared to both single-stage designs. CONCLUSION: Our comparative study demonstrates that the two-stage concept offers favorable hemolysis measures compared to single-stage designs, as the reduction in peak shear stresses outweighs the disadvantage associated with prolonged residence time.

Authors

Institutions

Publication Details

Journal
Artificial Organs
Published
2026-05-01
DOI
https://doi.org/10.1111/aor.70152
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts

Bernhard Semlitsch, Daniel Zimpfer, Bente Thamsen, Marcus Granegger et al.
Artificial Organs
Mechanical Circulatory Support Devices
article

Comparison of a Miniaturized Implantable Two‐Stage Rotodynamic Blood Pump to Equivalent Single‐Stage Concepts

Bernhard Semlitsch, Daniel Zimpfer, Bente Thamsen, Marcus Granegger, Sarah Linnemeier, Benjamin Torner, Stefan Maric
article en

Abstract

BACKGROUND: Hemocompatibility-related adverse events still limit the broader adoption of rotodynamic blood pumps. This study aims to evaluate the potential of a multistage pump design to enhance hemocompatibility in left ventricular assist devices. METHODS: Computational fluid dynamic simulations were utilized to compare the pump performance and hemocompatibility of a Two-Stage Pump to two equivalent single-stage devices, designed for identical operating conditions. One single-stage design features equal impeller and volute dimensions; the other, equal specific speed and specific diameter. Our simulation setup was validated against in vitro experiments of the pump performance of the Two-Stage Pump. RESULTS: The Two-Stage Pump achieved a substantial reduction in circumferential velocity (24% and 29% reduction). Although the two-stage configuration thereby effectively reduces shear stress, it simultaneously translates to a longer residence time due to the larger priming volume. However, because the numerical blood damage estimation weighs shear stress exponentially more than residence time, the beneficial effect of reduced shear stress partially compensates for the longer residence time. As a result, the Two-Stage Pump demonstrates a lower hemolysis index (12% and 14% reduction) compared to both single-stage designs. CONCLUSION: Our comparative study demonstrates that the two-stage concept offers favorable hemolysis measures compared to single-stage designs, as the reduction in peak shear stresses outweighs the disadvantage associated with prolonged residence time.

Artificial Organs
TU Wien (AT), Medical University of Vienna (AT)
Österreichische Forschungsförderungsgesellschaft
Openalex Percentile: Top 12%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.