Quantitative comparison of vessel deformation among four stent retrievers using a strain sensor equipped cerebrovascular model

BACKGROUND: Mechanical thrombectomy with stent retrievers is the standard treatment for acute large vessel occlusion stroke, but device induced vessel deformation during retrieval has not been quantified in a spatially resolved manner. METHODS: A polyvinyl alcohol (PVA) hydrogel cerebrovascular model reproducing the internal carotid artery to M1-M2 segments was equipped with eight strain sensors (TML FLA-03-11, gauge length 0.3 mm) at anatomically defined locations, corresponding to presumed perforator origins. Four stent retrievers (two sheet-type (Solitaire and pRESET) and two segmented-type (eNVi-SR and Embotrap)) were deployed and retrieved under identical conditions (three trials each). Primary outcomes were early peak strain and late mean strain. The Friedman test with post hoc Wilcoxon signed rank tests (Bonferroni correction) was used. RESULTS: The highest strain concentrations were consistently observed at the M1 outer curvature and distal perforator bearing M1 segment. Global early peak strain differed significantly among devices (P=0.025): Solitaire 56.7±5.3, pRESET 63.2±14.1, eNVi-SR 68.8±7.3, and Embotrap 84.9±1.9. Embotrap was significantly higher than Solitaire (P=0.037) and pRESET (P=0.019). For global late mean strain, Embotrap (29.3±2.9) was significantly higher than Solitaire (18.0±4.3; P=0.007). The relaxation ratio did not differ among devices (P=0.73). CONCLUSIONS: Strain sensor equipped PVA hydrogel models enabled reproducible, spatially resolved comparison of device induced vessel deformation. Embotrap induced significantly greater deformation than Solitaire and pRESET, particularly at perforator bearing segments.

Authors

Institutions

Publication Details

Journal
Journal of NeuroInterventional Surgery
Published
2026-04-30
DOI
https://doi.org/10.1136/jnis-2026-025413
Primary Topic
Cerebrovascular and Carotid Artery Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative comparison of vessel deformation among four stent retrievers using a strain sensor equipped cerebrovascular model

Kazunori Todoroki, Shunya Hanakita, Satoshi Iihoshi, Tomoko Ikemoto
Journal of NeuroInterventional Surgery
Cerebrovascular and Carotid Artery Diseases
article

Quantitative comparison of vessel deformation among four stent retrievers using a strain sensor equipped cerebrovascular model

Kazunori Todoroki, Shunya Hanakita, Satoshi Iihoshi, Tomoko Ikemoto
article en

Abstract

BACKGROUND: Mechanical thrombectomy with stent retrievers is the standard treatment for acute large vessel occlusion stroke, but device induced vessel deformation during retrieval has not been quantified in a spatially resolved manner. METHODS: A polyvinyl alcohol (PVA) hydrogel cerebrovascular model reproducing the internal carotid artery to M1-M2 segments was equipped with eight strain sensors (TML FLA-03-11, gauge length 0.3 mm) at anatomically defined locations, corresponding to presumed perforator origins. Four stent retrievers (two sheet-type (Solitaire and pRESET) and two segmented-type (eNVi-SR and Embotrap)) were deployed and retrieved under identical conditions (three trials each). Primary outcomes were early peak strain and late mean strain. The Friedman test with post hoc Wilcoxon signed rank tests (Bonferroni correction) was used. RESULTS: The highest strain concentrations were consistently observed at the M1 outer curvature and distal perforator bearing M1 segment. Global early peak strain differed significantly among devices (P=0.025): Solitaire 56.7±5.3, pRESET 63.2±14.1, eNVi-SR 68.8±7.3, and Embotrap 84.9±1.9. Embotrap was significantly higher than Solitaire (P=0.037) and pRESET (P=0.019). For global late mean strain, Embotrap (29.3±2.9) was significantly higher than Solitaire (18.0±4.3; P=0.007). The relaxation ratio did not differ among devices (P=0.73). CONCLUSIONS: Strain sensor equipped PVA hydrogel models enabled reproducible, spatially resolved comparison of device induced vessel deformation. Embotrap induced significantly greater deformation than Solitaire and pRESET, particularly at perforator bearing segments.

Journal of NeuroInterventional Surgery
Social Insurance Saitama Chuo Hospital (JP), Saitama Medical University (JP)
Openalex Percentile: Top 11%
Cerebrovascular and Carotid Artery Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.