Comparison of physical approaches to inter-crystal scattering in a staggered-layer scintillation PET detector: a simulation study

OBJECTIVE: Inter-crystal scattering (ICS) degrades spatial resolution and contrast by causing spurious lines of response in small-pitch, staggered three-layer pixelated LSO scintillation detectors used in a high-resolution BrainPET-7T insert. This work aimed to investigate the distribution of ICS events in our BrainPET-7T detector and understand its underlying mechanisms. APPROACH: Six physical approaches for computing probabilities to identify the first interaction positions of ICS events were studied using Geant4 simulations, focusing specifically on events with two interactions, which represent the most frequent ICS occurrence. Performance was evaluated using ideal interaction positions and energies directly from simulations, ground truth crystal-center positions with ideal energies, as well as crystal-center positions and energies determined by a maximum likelihood positioning (MLP) algorithm. MAIN RESULTS: A method based on Klein-Nishina probabilities, which were computed using geometric information (KN-G) and MLP-derived crystal-center positions and energies, achieved superior accuracy, particularly when the distance between successive interactions in Compton events was shorter than 6 mm and when incident angles were below 15 • . This advantage is partly attributable to crystal-center discretization, which masks a substantial portion of backscattering events. It reached a peak accuracy of 92.2 % at a distance of 19 mm. SIGNIFICANCE: These findings demonstrate the feasibility of combining the KN-G method with the MLP algorithm to mitigate the ICS effects, with the potential to improve the image quality of the 7T BrainPET insert.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-04-30
DOI
https://doi.org/10.1088/1361-6560/ae674b
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of physical approaches to inter-crystal scattering in a staggered-layer scintillation PET detector: a simulation study

Christoph Werner Lerche, Q Zhang, Cláudia Régio Brambilla, D. Niekämper et al.
Physics in Medicine and Biology
Radiation Detection and Scintillator Technologies
article

Comparison of physical approaches to inter-crystal scattering in a staggered-layer scintillation PET detector: a simulation study

Christoph Werner Lerche, Q Zhang, Cláudia Régio Brambilla, D. Niekämper, Qi Liu, Keith George Ciantar, Elisabeth Pfaehler, Nadim Jon Shah, J Scheins
article en

Abstract

OBJECTIVE: Inter-crystal scattering (ICS) degrades spatial resolution and contrast by causing spurious lines of response in small-pitch, staggered three-layer pixelated LSO scintillation detectors used in a high-resolution BrainPET-7T insert. This work aimed to investigate the distribution of ICS events in our BrainPET-7T detector and understand its underlying mechanisms. APPROACH: Six physical approaches for computing probabilities to identify the first interaction positions of ICS events were studied using Geant4 simulations, focusing specifically on events with two interactions, which represent the most frequent ICS occurrence. Performance was evaluated using ideal interaction positions and energies directly from simulations, ground truth crystal-center positions with ideal energies, as well as crystal-center positions and energies determined by a maximum likelihood positioning (MLP) algorithm. MAIN RESULTS: A method based on Klein-Nishina probabilities, which were computed using geometric information (KN-G) and MLP-derived crystal-center positions and energies, achieved superior accuracy, particularly when the distance between successive interactions in Compton events was shorter than 6 mm and when incident angles were below 15 • . This advantage is partly attributable to crystal-center discretization, which masks a substantial portion of backscattering events. It reached a peak accuracy of 92.2 % at a distance of 19 mm. SIGNIFICANCE: These findings demonstrate the feasibility of combining the KN-G method with the MLP algorithm to mitigate the ICS effects, with the potential to improve the image quality of the 7T BrainPET insert.

Physics in Medicine and Biology
Forschungszentrum Jülich (DE), Chengdu University of Technology (CN)
European Commission, Deutsche Forschungsgemeinschaft, National Natural Science Foundation of China
Openalex Percentile: Top 8%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.