Mitigating CT number variability between scanners, tube potentials, and patient sizes using spectral CT virtual monoenergetic imaging

OBJECTIVE: This study assessed CT number consistency in different materials and tissues as x-ray tube potential, scanner model and manufacturer, and phantom size were varied. We investigated using virtual monoenergetic imaging (VMIs) to improve CT number consistency compared to single-energy (SE) images. 
Approach: Four clinically-relevant material inserts (calcium, adipose, iodine, soft tissue) were placed into two phantom sizes (28-cm and 40-cm width) and scanned at each available tube potential on three CT scanners with distinct methods of spectral data collection. SE images and VMIs were reconstructed for each scan. The average CT number of each insert was calculated for each image series, and the consistency of this average value across different combinations of scan parameters was compared between VMIs and SE images using coefficient of variation (COV), median, range and a linear model estimating scanner-based CT number differences.
Main results: The COV for calcium, adipose, iodine and soft tissue inserts decreased from 10.1%, 9.7%, 28.8%, and 6.2% in SE images to 1.6%, 3.1%, 2.9%, and 5.7% in VMIs, respectively. When only tube potential was changed, the median standard deviation in insert average CT number decreased in VMIs compared to SE images in all materials. VMIs decreased the range of insert values between phantom sizes. VMIs decreased CT number variation due to scanner in both phantom sizes for iodine and in the calcium, 28-cm phantom. Inter-manufacturer differences in VMI CT numbers were observed in low density materials and the 40-cm phantom. 
Significance: Scan parameter variation affects CT numbers, negatively impacting quantitative imaging consistency and downstream patient metrics. This limits measurement standardization in serial studies, diverse cohorts, and between hospitals with data collection differences. VMIs from spectral CT can reduce this variability by providing more consistent material CT numbers relative to changes in tube potential, scanner model, and patient size.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-04-30
DOI
https://doi.org/10.1088/1361-6560/ae6749
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mitigating CT number variability between scanners, tube potentials, and patient sizes using spectral CT virtual monoenergetic imaging

Kishore Rajendran, Fang‐Shu Ou, Aeden Kayle Davis, Cynthia H McCollough et al.
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Mitigating CT number variability between scanners, tube potentials, and patient sizes using spectral CT virtual monoenergetic imaging

Kishore Rajendran, Fang‐Shu Ou, Aeden Kayle Davis, Cynthia H McCollough, Shuai Leng
article en

Abstract

OBJECTIVE: This study assessed CT number consistency in different materials and tissues as x-ray tube potential, scanner model and manufacturer, and phantom size were varied. We investigated using virtual monoenergetic imaging (VMIs) to improve CT number consistency compared to single-energy (SE) images. 
Approach: Four clinically-relevant material inserts (calcium, adipose, iodine, soft tissue) were placed into two phantom sizes (28-cm and 40-cm width) and scanned at each available tube potential on three CT scanners with distinct methods of spectral data collection. SE images and VMIs were reconstructed for each scan. The average CT number of each insert was calculated for each image series, and the consistency of this average value across different combinations of scan parameters was compared between VMIs and SE images using coefficient of variation (COV), median, range and a linear model estimating scanner-based CT number differences.
Main results: The COV for calcium, adipose, iodine and soft tissue inserts decreased from 10.1%, 9.7%, 28.8%, and 6.2% in SE images to 1.6%, 3.1%, 2.9%, and 5.7% in VMIs, respectively. When only tube potential was changed, the median standard deviation in insert average CT number decreased in VMIs compared to SE images in all materials. VMIs decreased the range of insert values between phantom sizes. VMIs decreased CT number variation due to scanner in both phantom sizes for iodine and in the calcium, 28-cm phantom. Inter-manufacturer differences in VMI CT numbers were observed in low density materials and the 40-cm phantom. 
Significance: Scan parameter variation affects CT numbers, negatively impacting quantitative imaging consistency and downstream patient metrics. This limits measurement standardization in serial studies, diverse cohorts, and between hospitals with data collection differences. VMIs from spectral CT can reduce this variability by providing more consistent material CT numbers relative to changes in tube potential, scanner model, and patient size.

Physics in Medicine and Biology
Mayo Clinic (US)
Siemens Healthineers, National Institute of Biomedical Imaging and Bioengineering
Openalex Percentile: Top 17%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.