Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial

Aim: To evaluate whether, in children treated in paediatric intensive care units (PICU) with peripheral venous access devices (PVADs) receiving intermediate‐ or high‐risk peripheral intravenous medications, the use of a tissue‐infiltration biosensor (ivWatch), compared to standard care, reduces extravasation injuries. Methods: A superiority, parallel group randomised controlled trial conducted in a single Queensland quaternary hospital PICU from September 2020 to July 2022. Children with PVADs inserted in the distal half of the limb and receiving intermediate‐ to high‐risk infusates for ≥ 24 h were randomly allocated in a 1:1 ratio to biosensor (ivWatch monitor applied within 10–15 mm of the PVAD tip) plus standard clinical observation, or standard observation alone. Primary outcome was extravasation severity on the 4‐point Cincinnati Children's Hospital Medical Centre Extravasation Harm Scale. Injuries were compared between treatment groups using ordinal logistic regression. Results: A total of 175 children were enrolled, 88 to biosensor and 87 to standard care. Extravasation severity was similar between groups, with no, mild and moderate injury occurring in 56 (69%), 25 (30%) and 2 (2%) participants in the biosensor group and 61 (70%), 24 (28%) and 2 (2%) in the standard care group (proportional odds ratio = 1.13; 95% CI: 0.59, 2.15, p = 0.71). No statistically significant between‐group differences in extravasation severity, volume, treatment sequelae, PVADs used or dwell time were observed. The biosensor was acceptable to nurses. Conclusions: In this PICU population, biosensor monitoring in addition to standard care did not significantly reduce extravasation injuries. The technology was acceptable to bedside nursing staff. Trial Registration: ACTRN12620000317998

Authors

Publication Details

Journal
Universität Zürich, ZORA
Published
2026-04-14
DOI
https://doi.org/10.5167/uzh-433843
Primary Topic
Chemotherapy-related skin toxicity
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial

Tricia M. Kleidon, Sarfaraz Rahiman, Robert S. Ware, Rebecca Paterson et al.
Universität Zürich, ZORA
Chemotherapy-related skin toxicity
article

Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial

Tricia M. Kleidon, Sarfaraz Rahiman, Robert S. Ware, Rebecca Paterson, Joshua Byrnes, Craig A. McBride, Jessica A. Schults, Thim Baveas, Melanie Kennedy, Luregn Schlapbach, Amanda J. Ullman
article en

Abstract

Aim: To evaluate whether, in children treated in paediatric intensive care units (PICU) with peripheral venous access devices (PVADs) receiving intermediate‐ or high‐risk peripheral intravenous medications, the use of a tissue‐infiltration biosensor (ivWatch), compared to standard care, reduces extravasation injuries. Methods: A superiority, parallel group randomised controlled trial conducted in a single Queensland quaternary hospital PICU from September 2020 to July 2022. Children with PVADs inserted in the distal half of the limb and receiving intermediate‐ to high‐risk infusates for ≥ 24 h were randomly allocated in a 1:1 ratio to biosensor (ivWatch monitor applied within 10–15 mm of the PVAD tip) plus standard clinical observation, or standard observation alone. Primary outcome was extravasation severity on the 4‐point Cincinnati Children's Hospital Medical Centre Extravasation Harm Scale. Injuries were compared between treatment groups using ordinal logistic regression. Results: A total of 175 children were enrolled, 88 to biosensor and 87 to standard care. Extravasation severity was similar between groups, with no, mild and moderate injury occurring in 56 (69%), 25 (30%) and 2 (2%) participants in the biosensor group and 61 (70%), 24 (28%) and 2 (2%) in the standard care group (proportional odds ratio = 1.13; 95% CI: 0.59, 2.15, p = 0.71). No statistically significant between‐group differences in extravasation severity, volume, treatment sequelae, PVADs used or dwell time were observed. The biosensor was acceptable to nurses. Conclusions: In this PICU population, biosensor monitoring in addition to standard care did not significantly reduce extravasation injuries. The technology was acceptable to bedside nursing staff. Trial Registration: ACTRN12620000317998

Universität Zürich, ZORA
Good health and well-being
Openalex Percentile: Top 5%
Chemotherapy-related skin toxicity
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.