Performance of a Vendor System for Systematic Offline Adaptive Breast Helical Radiotherapy

Background: This study evaluated the performance of a commercial offline adaptive radiotherapy system for systematic monitoring of breast cancer treatment with nodal irradiation using helical tomotherapy. Methods: Thirty patients treated for invasive unilateral breast carcinoma were analysed. For each patient, three megavoltage CT scans acquired at the first, middle, and last treatment sessions were processed through the PreciseART (Accuray, US) offline ART workflow. Automatically deformed structures were compared with manually delineated reference structures. Geometric accuracy was assessed using the Dice similarity coefficient (DSC), Hausdorff distance (HD95), mean distance to agreement (MDA), and barycentre distance (BD). The dosimetric parameters included D2% and V95% for targets and Dmean/Dmax/V20Gy for organs at risk. Results: Median DSCs exceeded 0.9 for the CTVbreast, PTVbreast, heart, and ipsilateral lung and were above 0.8 for the remaining structures, except the CTVn and oesophagus. Dosimetric differences between deformed and reference structures were within 5% for D2% across all targets and for V95% of the CTVbreast and PTVbreast in 90% of the sessions. The ipsilateral lung V20Gy differed by less than 5% in more than 90% of the sessions. Larger deviations (up to 10%) were observed for the nodal PTVs and mean heart dose, while the greatest inconsistencies were found for the oesophagus and spinal canal. Conclusions: The evaluated offline ART system demonstrates sufficient accuracy for automated monitoring of breast and lung structures. However, cautious interpretation remains necessary for nodal targets, heart, and oesophagus dosimetry prior to clinical implementation.

Authors

Institutions

Publication Details

Journal
Cancers
Published
2026-04-27
DOI
https://doi.org/10.3390/cancers18091386
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of a Vendor System for Systematic Offline Adaptive Breast Helical Radiotherapy

P. Meyer, C. Niederst, Georges Noël, Florence Arbor et al.
Cancers
Advanced Radiotherapy Techniques
article

Performance of a Vendor System for Systematic Offline Adaptive Breast Helical Radiotherapy

P. Meyer, C. Niederst, Georges Noël, Florence Arbor, Claire Dossun, Loris Barrier, Anthony Richert
article en

Abstract

Background: This study evaluated the performance of a commercial offline adaptive radiotherapy system for systematic monitoring of breast cancer treatment with nodal irradiation using helical tomotherapy. Methods: Thirty patients treated for invasive unilateral breast carcinoma were analysed. For each patient, three megavoltage CT scans acquired at the first, middle, and last treatment sessions were processed through the PreciseART (Accuray, US) offline ART workflow. Automatically deformed structures were compared with manually delineated reference structures. Geometric accuracy was assessed using the Dice similarity coefficient (DSC), Hausdorff distance (HD95), mean distance to agreement (MDA), and barycentre distance (BD). The dosimetric parameters included D2% and V95% for targets and Dmean/Dmax/V20Gy for organs at risk. Results: Median DSCs exceeded 0.9 for the CTVbreast, PTVbreast, heart, and ipsilateral lung and were above 0.8 for the remaining structures, except the CTVn and oesophagus. Dosimetric differences between deformed and reference structures were within 5% for D2% across all targets and for V95% of the CTVbreast and PTVbreast in 90% of the sessions. The ipsilateral lung V20Gy differed by less than 5% in more than 90% of the sessions. Larger deviations (up to 10%) were observed for the nodal PTVs and mean heart dose, while the greatest inconsistencies were found for the oesophagus and spinal canal. Conclusions: The evaluated offline ART system demonstrates sufficient accuracy for automated monitoring of breast and lung structures. However, cautious interpretation remains necessary for nodal targets, heart, and oesophagus dosimetry prior to clinical implementation.

CancersVol. 18(9)
Laboratoire des Sciences de l'Ingénieur, de l'Informatique et de l'Imagerie (FR), Levi Strauss (United States) (US), Hôpital Civil, Strasbourg (FR)
Good health and well-being
Openalex Percentile: Top 5%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.