Temporary Loss of Pacemaker Capture During Focal Pulsed Field Ablation: A Case of Reversible Lead–Energy Interactions

BACKGROUND: Pulsed field ablation (PFA) is increasingly used for atrial arrhythmia ablation and is considered safe in patients with cardiac implantable electronic devices, though intraprocedural lead interactions remain undercharacterized. CASE: We report a 75-year-old woman with a dual-chamber pacemaker who underwent focal PFA for recurrent atrial tachycardia using a 9-mm lattice-tip catheter. Immediately following PFA near the superior vena cava-right atrial junction, atrial lead capture was lost with transient right atrial lead threshold elevation and T-wave oversensing. All parameters normalized within 1 h. CONCLUSIONS: This case highlights reversible lead-energy interactions and the importance of real-time device monitoring when ablating near indwelling leads.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Electrophysiology
Published
2026-04-26
DOI
https://doi.org/10.1111/jce.70308
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Temporary Loss of Pacemaker Capture During Focal Pulsed Field Ablation: A Case of Reversible Lead–Energy Interactions

Rahul Devathu, Tina Baykaner, Kushal Chatterjee, Tasmeen Hussain et al.
Journal of Cardiovascular Electrophysiology
Cardiac Arrhythmias and Treatments
article

Temporary Loss of Pacemaker Capture During Focal Pulsed Field Ablation: A Case of Reversible Lead–Energy Interactions

Rahul Devathu, Tina Baykaner, Kushal Chatterjee, Tasmeen Hussain, Samhith Kambampati, Daniel Joseph Gonzalez, Aaryamaan Verma, Muhammad Fazal, Erick Godinez
article en

Abstract

BACKGROUND: Pulsed field ablation (PFA) is increasingly used for atrial arrhythmia ablation and is considered safe in patients with cardiac implantable electronic devices, though intraprocedural lead interactions remain undercharacterized. CASE: We report a 75-year-old woman with a dual-chamber pacemaker who underwent focal PFA for recurrent atrial tachycardia using a 9-mm lattice-tip catheter. Immediately following PFA near the superior vena cava-right atrial junction, atrial lead capture was lost with transient right atrial lead threshold elevation and T-wave oversensing. All parameters normalized within 1 h. CONCLUSIONS: This case highlights reversible lead-energy interactions and the importance of real-time device monitoring when ablating near indwelling leads.

Journal of Cardiovascular Electrophysiology
University of Nevada, Reno (US), Stanford University (US)
Affordable and clean energy
Openalex Percentile: Top 5%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.