Agreement of wearable and standing bioimpedance devices with bioimpedance spectroscopy estimates of total body water in adults with Down syndrome

Abstract Background Wearable bioimpedance technology offers a potentially accessible method for assessing total body water (TBW), yet its agreement with laboratory-based approaches in adults with Down syndrome remains unclear. The purpose of this study was to evaluate whether a wearable bioimpedance device demonstrates agreement with laboratory-based bioelectrical impedance spectroscopy (BIS) estimates of TBW comparable to standing bioimpedance devices in adults with Down syndrome. Methods Sixteen adults with Down syndrome (age: 28.3 ± 9.2 year; height: 150.9 ± 7.8 cm; body mass: 77.4 ± 20.3 kg) completed TBW assessment using BIS (reference), a wearable bioimpedance device (BIA W ), and two standing multi-frequency bioimpedance analyzers (BIA ST1 , BIA ST2 ). Results Repeated measures ANOVA revealed significant differences among devices (F (3,42) = 17.52, p < 0.001). Both standing devices produced lower TBW values compared with BIS (p < 0.001; d = 1.32 and 1.00), whereas BIA W did not differ from BIS (p = 0.106; d = 0.45). All devices demonstrated strong correlations with BIS (r = 0.92–0.93). Bland–Altman analyses showed systematic underestimation for standing devices (− 3.12 ± 4.65 L and − 2.36 ± 4.64 L), while the wearable exhibited the smallest bias (− 1.07 ± 4.64 L). Standard error of estimate values were similar across devices (2.34–2.45 L). Conclusion These findings indicate that wearable bioimpedance technology demonstrated closer agreement with laboratory-based TBW estimates than standing devices.

Authors

Institutions

Publication Details

Journal
European Journal of Clinical Nutrition
Published
2026-11-09
DOI
https://doi.org/10.1038/s41430-026-01848-6
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Agreement of wearable and standing bioimpedance devices with bioimpedance spectroscopy estimates of total body water in adults with Down syndrome

Brett Nickerson, Michael R. Esco, Elroy J. Aguiar, Kristen Nichols Heitman et al.
1 links
European Journal of Clinical Nutrition
Body Composition Measurement Techniques
article

Agreement of wearable and standing bioimpedance devices with bioimpedance spectroscopy estimates of total body water in adults with Down syndrome

Brett Nickerson, Michael R. Esco, Elroy J. Aguiar, Kristen Nichols Heitman, Cara Chase, Stefan A. Czerwinski, Matthew Ali Mohammadnabi, Justo Perez, Jeff Pan
article en

Abstract

Abstract Background Wearable bioimpedance technology offers a potentially accessible method for assessing total body water (TBW), yet its agreement with laboratory-based approaches in adults with Down syndrome remains unclear. The purpose of this study was to evaluate whether a wearable bioimpedance device demonstrates agreement with laboratory-based bioelectrical impedance spectroscopy (BIS) estimates of TBW comparable to standing bioimpedance devices in adults with Down syndrome. Methods Sixteen adults with Down syndrome (age: 28.3 ± 9.2 year; height: 150.9 ± 7.8 cm; body mass: 77.4 ± 20.3 kg) completed TBW assessment using BIS (reference), a wearable bioimpedance device (BIA W ), and two standing multi-frequency bioimpedance analyzers (BIA ST1 , BIA ST2 ). Results Repeated measures ANOVA revealed significant differences among devices (F (3,42) = 17.52, p < 0.001). Both standing devices produced lower TBW values compared with BIS (p < 0.001; d = 1.32 and 1.00), whereas BIA W did not differ from BIS (p = 0.106; d = 0.45). All devices demonstrated strong correlations with BIS (r = 0.92–0.93). Bland–Altman analyses showed systematic underestimation for standing devices (− 3.12 ± 4.65 L and − 2.36 ± 4.64 L), while the wearable exhibited the smallest bias (− 1.07 ± 4.64 L). Standard error of estimate values were similar across devices (2.34–2.45 L). Conclusion These findings indicate that wearable bioimpedance technology demonstrated closer agreement with laboratory-based TBW estimates than standing devices.

European Journal of Clinical Nutrition
University of Alabama (US), The Ohio State University (US)
Good health and well-being
Body Composition Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon