The impact of DICOM import/export on radiotherapy structures in commercial systems

BACKGROUND AND PURPOSE: To investigate the impact that loading contours into an internal representation and saving them back to DICOM format has on contour position within commercially available contouring systems. MATERIALS AND METHODS: Two DICOM RT Structure Sets (RTSS) were loaded into 12 different commercial contour editing systems. Small non-editing actions were taken to prompt the systems to convert the structures to their internal representation. Contours were exported again as DICOM RTSS. The exported contours were compared to the original contours. The impact on contour similarity measures was quantified by their calculation against a common reference set for both the original and the re-exported contours. RESULTS: Average distances up to 0.23 mm were observed between the original and re-exported contours on thoracic data for some systems, with maximum distances up to 37.7 mm resulting from topological changes. Contour boundary changes impacted contouring similarity measures up to 2.6%, 29%, 35% and 21% for volumetric Dice Similarity Coefficient, 2D Mean Surface Distance, 3D Mean Surface Distance, and Hausdorff Distance, respectively. Notably, Surface Dice Similarity Coefficient and Normalised Added Path Length varied up to 96% and 97% respectively at low tolerances. Other systems preserved the original contours. CONCLUSIONS: Caution should be employed when comparing contouring accuracy measures between studies or within multi-centre evaluations if different contouring systems have been used. For monitoring auto-contouring in clinical practice, differences between automatic and clinical contours smaller than one quarter of the in-plane voxel size can be disregarded, as possibly reflecting software-related effects rather than intentional editing.

Authors

Institutions

Publication Details

Journal
Physics and Imaging in Radiation Oncology
Published
2026-04-26
DOI
https://doi.org/10.1016/j.phro.2026.100981
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of DICOM import/export on radiotherapy structures in commercial systems

Ciaran Malone, Daniel Rossiter, Djamal Boukerroui, Georgios Tsekas et al.
Physics and Imaging in Radiation Oncology
Advanced Radiotherapy Techniques
article

The impact of DICOM import/export on radiotherapy structures in commercial systems

Ciaran Malone, Daniel Rossiter, Djamal Boukerroui, Georgios Tsekas, Mark J. Gooding, Christina Skourou, Kieran Venner, Annamieke Koops, Maxwell Robinson, David Nash, Rita Simões
article en

Abstract

BACKGROUND AND PURPOSE: To investigate the impact that loading contours into an internal representation and saving them back to DICOM format has on contour position within commercially available contouring systems. MATERIALS AND METHODS: Two DICOM RT Structure Sets (RTSS) were loaded into 12 different commercial contour editing systems. Small non-editing actions were taken to prompt the systems to convert the structures to their internal representation. Contours were exported again as DICOM RTSS. The exported contours were compared to the original contours. The impact on contour similarity measures was quantified by their calculation against a common reference set for both the original and the re-exported contours. RESULTS: Average distances up to 0.23 mm were observed between the original and re-exported contours on thoracic data for some systems, with maximum distances up to 37.7 mm resulting from topological changes. Contour boundary changes impacted contouring similarity measures up to 2.6%, 29%, 35% and 21% for volumetric Dice Similarity Coefficient, 2D Mean Surface Distance, 3D Mean Surface Distance, and Hausdorff Distance, respectively. Notably, Surface Dice Similarity Coefficient and Normalised Added Path Length varied up to 96% and 97% respectively at low tolerances. Other systems preserved the original contours. CONCLUSIONS: Caution should be employed when comparing contouring accuracy measures between studies or within multi-centre evaluations if different contouring systems have been used. For monitoring auto-contouring in clinical practice, differences between automatic and clinical contours smaller than one quarter of the in-plane voxel size can be disregarded, as possibly reflecting software-related effects rather than intentional editing.

Physics and Imaging in Radiation OncologyVol. 39
University Medical Center Groningen (NL), University of Manchester (GB), The Netherlands Cancer Institute (NL), Portsmouth Hospitals NHS Trust (GB), St. Luke's Hospital (IE), Oxford University Hospitals NHS Trust (GB), Evotec (United Kingdom) (GB), Erasmus MC Cancer Institute (NL)
Openalex Percentile: Top 14%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.