Determination of a safe radiation distance from a 16-slice multi-detector computed tomography (MDCT) scan machine

The objective of this study is to determine the location with the lowest radiation levels inside the CT scan room while performing examinations using a 16-slice multi-detector computed tomography (MDCT) scanner. This was an analytical, experimental study taking radiation dose measurements using a dosimeter placed one location at a time using a clock-face pattern with 1-m intervals during a simulated CT-guided biopsy procedure. A total of 51 dose measurements were taken. The lowest radiation dose at 1 m is at the 9:00 position, then at the 3:00 position. The lowest radiation dose at 2 m is at the 9:00 position. The doses in these three positions do not have a statistical difference, but are significantly lower than the rest of the doses in other positions. The dose of radiation in a CT scan room is dependent on the distance and direction of the radiation source. If medical personnel need to stay inside the room during procedures, the safer distance from the radiation source was 1 and 2 m away at the 9:00 position and 1 m away at the 3:00 position, which are on the sides of the CT scanner.

Authors

Institutions

Publication Details

Journal
Annals of the ICRP
Published
2026-04-24
DOI
https://doi.org/10.1177/01466453251412144
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of a safe radiation distance from a 16-slice multi-detector computed tomography (MDCT) scan machine

A.I. Hocson, N.D.P. Concepcion
Annals of the ICRP
Radiation Dose and Imaging
article

Determination of a safe radiation distance from a 16-slice multi-detector computed tomography (MDCT) scan machine

A.I. Hocson, N.D.P. Concepcion
article en

Abstract

The objective of this study is to determine the location with the lowest radiation levels inside the CT scan room while performing examinations using a 16-slice multi-detector computed tomography (MDCT) scanner. This was an analytical, experimental study taking radiation dose measurements using a dosimeter placed one location at a time using a clock-face pattern with 1-m intervals during a simulated CT-guided biopsy procedure. A total of 51 dose measurements were taken. The lowest radiation dose at 1 m is at the 9:00 position, then at the 3:00 position. The lowest radiation dose at 2 m is at the 9:00 position. The doses in these three positions do not have a statistical difference, but are significantly lower than the rest of the doses in other positions. The dose of radiation in a CT scan room is dependent on the distance and direction of the radiation source. If medical personnel need to stay inside the room during procedures, the safer distance from the radiation source was 1 and 2 m away at the 9:00 position and 1 m away at the 3:00 position, which are on the sides of the CT scanner.

Annals of the ICRP
St. Luke's Medical Center (PH)
Good health and well-being
Openalex Percentile: Top 4%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.