Simultaneous determination of monochloramine and common anions in drinking water by mixed-mode ion chromatography with sequential UV absorbance and suppressed conductivity detection

A sensitive and rapid mixed-mode ion chromatography-based method for the simultaneous quantification of the disinfectant monochloramine and common inorganic anions within treated drinking waters is presented. The separation was achieved on a high-capacity anion-exchange column (Dionex IonPac AS15, 4 mm x 250 mm) using a mixed-mode approach, which combined ion-exchange and ion-exclusion-based retention for the inorganic anions and monochloramine, respectively. An electrolytically produced potassium hydroxide eluent gradient was employed, with the method combining suppressed conductivity detection with in-house assembled 50-mm flow cell high-sensitivity UV absorbance detection. This configuration provides a five-fold enhancement in UV absorbance signal relative to a conventional 10-mm flow cell, substantially improving sensitivity for UV-absorbing species. Monochloramine is detected directly and without derivatisation, with an absorbance maximum at 240 nm. The optimised method was able to separate and quantify eight target analytes commonly monitored in treated drinking waters, namely fluoride, chloride, nitrite, bromide, nitrate, phosphate, sulphate, and monochloramine, all within a 13-minute runtime. The method was validated, demonstrating excellent linearity (r≥0.996) across a 7-point calibration range and high precision with %RSD consistently at or below 1.8 %. The LOD for monochloramine is 0.11 mg/L, while those for the other anions ranged from 1.5 µg/L for fluoride to 0.17 mg/L for nitrate. The method's limits of detection were well below the maximum levels recommended in World Health Organization (WHO) guidelines for drinking water. Notably, the system was selective for monochloramine detection in the presence of dichloramine.

Authors

Institutions

Publication Details

Journal
Journal of Chromatography A
Published
2026-04-24
DOI
https://doi.org/10.1016/j.chroma.2026.467031
Primary Topic
Water Treatment and Disinfection
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous determination of monochloramine and common anions in drinking water by mixed-mode ion chromatography with sequential UV absorbance and suppressed conductivity detection

Ibraam E. Mikhail, Robert A. Shellie, Brett Paull, Nanda Mrabawani
Journal of Chromatography A
Water Treatment and Disinfection
article

Simultaneous determination of monochloramine and common anions in drinking water by mixed-mode ion chromatography with sequential UV absorbance and suppressed conductivity detection

Ibraam E. Mikhail, Robert A. Shellie, Brett Paull, Nanda Mrabawani
article en

Abstract

A sensitive and rapid mixed-mode ion chromatography-based method for the simultaneous quantification of the disinfectant monochloramine and common inorganic anions within treated drinking waters is presented. The separation was achieved on a high-capacity anion-exchange column (Dionex IonPac AS15, 4 mm x 250 mm) using a mixed-mode approach, which combined ion-exchange and ion-exclusion-based retention for the inorganic anions and monochloramine, respectively. An electrolytically produced potassium hydroxide eluent gradient was employed, with the method combining suppressed conductivity detection with in-house assembled 50-mm flow cell high-sensitivity UV absorbance detection. This configuration provides a five-fold enhancement in UV absorbance signal relative to a conventional 10-mm flow cell, substantially improving sensitivity for UV-absorbing species. Monochloramine is detected directly and without derivatisation, with an absorbance maximum at 240 nm. The optimised method was able to separate and quantify eight target analytes commonly monitored in treated drinking waters, namely fluoride, chloride, nitrite, bromide, nitrate, phosphate, sulphate, and monochloramine, all within a 13-minute runtime. The method was validated, demonstrating excellent linearity (r≥0.996) across a 7-point calibration range and high precision with %RSD consistently at or below 1.8 %. The LOD for monochloramine is 0.11 mg/L, while those for the other anions ranged from 1.5 µg/L for fluoride to 0.17 mg/L for nitrate. The method's limits of detection were well below the maximum levels recommended in World Health Organization (WHO) guidelines for drinking water. Notably, the system was selective for monochloramine detection in the presence of dichloramine.

Journal of Chromatography AVol. 1779
University of Tasmania (AU)
Australian Research Council
Clean water and sanitation
Openalex Percentile: Top 16%
Water Treatment and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.