Optimizing Reconstruction Parameters for Detecting Peripheral In-Stent Restenosis with Photon-Counting Detector CT: A Phantom Study

Background/Objectives: To determine the optimal reconstruction parameters for accurate visualization of peripheral in-stent restenosis using photon-counting detector CT (PCD-CT), and to evaluate its potential advantages over energy-integrated detector CT (EID-CT). Methods: Endovascular peripheral stents with varying degrees of in-stent restenosis were scanned in a custom-made phantom using EID-CT (Somatom Force) and PCD-CT (Naeotom Alpha) under clinical acquisition protocols. EID-CT images were reconstructed with Bv40 and Bv59 kernels at 512 matrices. PCD-CT data were acquired in standard-resolution (SR) and ultra-high-resolution (UHR) modes. In both modes, images were reconstructed with multiple kernels (Bv40, Bv56 and Bv72) and matrix sizes (512 and 1024 matrix). In SR mode, additional virtual monoenergetic images (40–100 keV) were generated, while UHR mode included only polychromatic reconstructions. Quantitative image quality (noise, contrast, contrast-to-noise ratio [CNR]) was measured, and two blinded readers performed qualitative assessments of restenosis visualization. Results: PCD-CT with SR mode at VMI 40 keV achieved the highest image contrast and CNR, significantly outperforming EID-CT and PCD-CTUHR under matched conditions (all p < 0.05). The sharper reconstruction kernel further enhanced the image contrast and improved subjective visualization despite increased image noise. Both readers ranked PCD-CTSR-Bv72-40keV at 1024 matrix highest for detecting all degrees of restenosis, with excellent inter-reader agreement (ρ > 0.80). Conclusions: PCD-CT in SR mode at VMI 40 keV, specifically using the Bv72 kernel with a 1024 matrix, optimizes the visualization of peripheral in-stent restenosis. Compared to EID-CT, PCD-CT provides superior image quality and detectability of restenosis.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-04-22
DOI
https://doi.org/10.3390/diagnostics16091253
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing Reconstruction Parameters for Detecting Peripheral In-Stent Restenosis with Photon-Counting Detector CT: A Phantom Study

Reinoud P.H. Bokkers, Xinlin Zheng, Jean‐Paul P.M. de Vries, Joost F. Hop et al.
Diagnostics
Advanced X-ray and CT Imaging
article

Optimizing Reconstruction Parameters for Detecting Peripheral In-Stent Restenosis with Photon-Counting Detector CT: A Phantom Study

Reinoud P.H. Bokkers, Xinlin Zheng, Jean‐Paul P.M. de Vries, Joost F. Hop, Marcel J.W. Greuter, Yiheng Tan, Evie J. I. Hoeijmakers, Magdalena Dobrolinska
article en

Abstract

Background/Objectives: To determine the optimal reconstruction parameters for accurate visualization of peripheral in-stent restenosis using photon-counting detector CT (PCD-CT), and to evaluate its potential advantages over energy-integrated detector CT (EID-CT). Methods: Endovascular peripheral stents with varying degrees of in-stent restenosis were scanned in a custom-made phantom using EID-CT (Somatom Force) and PCD-CT (Naeotom Alpha) under clinical acquisition protocols. EID-CT images were reconstructed with Bv40 and Bv59 kernels at 512 matrices. PCD-CT data were acquired in standard-resolution (SR) and ultra-high-resolution (UHR) modes. In both modes, images were reconstructed with multiple kernels (Bv40, Bv56 and Bv72) and matrix sizes (512 and 1024 matrix). In SR mode, additional virtual monoenergetic images (40–100 keV) were generated, while UHR mode included only polychromatic reconstructions. Quantitative image quality (noise, contrast, contrast-to-noise ratio [CNR]) was measured, and two blinded readers performed qualitative assessments of restenosis visualization. Results: PCD-CT with SR mode at VMI 40 keV achieved the highest image contrast and CNR, significantly outperforming EID-CT and PCD-CTUHR under matched conditions (all p < 0.05). The sharper reconstruction kernel further enhanced the image contrast and improved subjective visualization despite increased image noise. Both readers ranked PCD-CTSR-Bv72-40keV at 1024 matrix highest for detecting all degrees of restenosis, with excellent inter-reader agreement (ρ > 0.80). Conclusions: PCD-CT in SR mode at VMI 40 keV, specifically using the Bv72 kernel with a 1024 matrix, optimizes the visualization of peripheral in-stent restenosis. Compared to EID-CT, PCD-CT provides superior image quality and detectability of restenosis.

DiagnosticsVol. 16(9)
University Medical Center Groningen (NL), Medical University of Silesia (PL), University of Groningen (NL), Maastricht University Medical Centre (NL), Maastricht University (NL)
Openalex Percentile: Top 12%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.