Background: Insertable loop recorder (ILR) services are increasingly constrained by high volumes of transmitted episodes, many of which are false, clinically irrelevant, or non-actionable. Cloud-based artificial intelligence (AI) algorithms have the potential to suppress false atrial fibrillation (AF) and pause alerts while preserving clinically relevant events. We present the first real-world impact of an AI algorithm activated across a fixed multicentre ILR cohort. Methods: We performed a retrospective, multicentre before-and-after cohort analysis of consecutive patients. Eligible patients had continuous ILR monitoring for 12 months before and 12 months after service-wide activation ("switch-on") of the Medtronic AccuRhythm AI platform (Medtronic plc, Galway, Ireland), enabling within-patient comparison. All clinician-facing transmitted alerts were counted in each period. Secondary analyses examined the concentration of alert burden across patients and estimated workflow impact using published time-and-motion data for remote transmission review. Differences in alert counts were tested using paired t-tests, reporting mean paired differences with 95% confidence intervals (CI). Results: The cohort included 445 patients (Reveal LINQ n=438; LINQ II n=7); 440 (99%) were implanted for syncope (mean age 67±15 years; 49.6% male). Total transmitted alert volume fell from 4,261 pre-AI to 2,509 post-AI (29% reduction). The mean paired within-patient change was −3.94 alerts (95% CI −7.5 to −0.4; p<0.05). Alerts were highly concentrated: 7% of patients generated >90% of alerts, and 104 patients had intermittent disconnection from remote monitoring. Applying established workflow timings (11-13 minutes per remote transmission review) translated the reduction into 185-218 hours of physiologist review time released annually: approximately four hours per week of "virtual physiologist" capacity. Conclusions: In routine practice, AI activation in a multicentre observational study was associated with a statistically significant reduction in ILR alert burden and a clinically meaningful release of staff capacity. Parallel management of high-alert patients and connectivity optimisation may further amplify the operational benefit.

Authors

Publication Details

Journal
Cureus
Published
2026-04-22
DOI
https://doi.org/10.7759/cureus.107519
Primary Topic
Cardiovascular Syncope and Autonomic Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Use of an Artificial Intelligence Algorithm to Increase Productivity in Implantable Loop Recorder Monitoring: A Multicentre Observational Study

Cherry Alexander, Sophie Bagnall, Faheem Ahmad, Ken Teh et al.
Cureus
Cardiovascular Syncope and Autonomic Disorders
article

Use of an Artificial Intelligence Algorithm to Increase Productivity in Implantable Loop Recorder Monitoring: A Multicentre Observational Study

Cherry Alexander, Sophie Bagnall, Faheem Ahmad, Ken Teh, Catherine Vaughan, Alan Robertson, Andrew Saunders
article en

Abstract

Background: Insertable loop recorder (ILR) services are increasingly constrained by high volumes of transmitted episodes, many of which are false, clinically irrelevant, or non-actionable. Cloud-based artificial intelligence (AI) algorithms have the potential to suppress false atrial fibrillation (AF) and pause alerts while preserving clinically relevant events. We present the first real-world impact of an AI algorithm activated across a fixed multicentre ILR cohort. Methods: We performed a retrospective, multicentre before-and-after cohort analysis of consecutive patients. Eligible patients had continuous ILR monitoring for 12 months before and 12 months after service-wide activation ("switch-on") of the Medtronic AccuRhythm AI platform (Medtronic plc, Galway, Ireland), enabling within-patient comparison. All clinician-facing transmitted alerts were counted in each period. Secondary analyses examined the concentration of alert burden across patients and estimated workflow impact using published time-and-motion data for remote transmission review. Differences in alert counts were tested using paired t-tests, reporting mean paired differences with 95% confidence intervals (CI). Results: The cohort included 445 patients (Reveal LINQ n=438; LINQ II n=7); 440 (99%) were implanted for syncope (mean age 67±15 years; 49.6% male). Total transmitted alert volume fell from 4,261 pre-AI to 2,509 post-AI (29% reduction). The mean paired within-patient change was −3.94 alerts (95% CI −7.5 to −0.4; p<0.05). Alerts were highly concentrated: 7% of patients generated >90% of alerts, and 104 patients had intermittent disconnection from remote monitoring. Applying established workflow timings (11-13 minutes per remote transmission review) translated the reduction into 185-218 hours of physiologist review time released annually: approximately four hours per week of "virtual physiologist" capacity. Conclusions: In routine practice, AI activation in a multicentre observational study was associated with a statistically significant reduction in ILR alert burden and a clinically meaningful release of staff capacity. Parallel management of high-alert patients and connectivity optimisation may further amplify the operational benefit.

Cureus
Decent work and economic growth
Openalex Percentile: Top 6%
Cardiovascular Syncope and Autonomic Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.